User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
122
SYSTem:BUZZer:FTIMe
Set
Query
Description
Sets the FAIL Sound duration in seconds.
Syntax
Query Syntax
SYSTem:BUZZer:FTIMe <NR2>
SYSTem:BUZZer:FTIMe?
Parameter/
Return parameter
<NR2>
0.2~999.9
Example
SYST:BUZZ:FTIM 1
Sets the buzzer to 1 second for a FAIL judgment.
SYSTem:ERRor
Query
Description
Returns any errors in the output buffer. See the
error code table below for details.
Query Syntax
SYSTem:ERRor ?
Return parameter
<string>
Returns an error string that includes
an error code and an error
description.
Error Code Table
Error code, Error description
0,No Error
20,Command Error
21,Volume Error
22,String Error
23,Query Error
24,Mode Error
25,Time Error
26,DC Over 50W
27,GBV > 5.4V
30,Voltage Setting Error
31,Current Setting Error
32,Current HI SET Error
33,Current LOW SET Error
34,Resistance HI SET Error
35,Resistance HI SET Error