User Manual

OPERATION
93
4. To stop the test when HOLD is
displayed on-screen, press the
STOP button.
STOP
Note
Stop a Running
Test
1. To stop the AUTO test at any time
when it is running, press the STOP
button. The AUTO test will stop
immediately. When the STOP
button is pressed, a judgment is
not made on the current test and
any remaining tests are aborted.
All panel keys except the STOP
and START buttons are locked
when the tester has been stopped.
All the results up until when the
AUTO test was stopped are shown
on-screen. See page 94 for more
details on automatic test results.
STOP
E MA U OT N A_ *A U T O = 0 0 1 - * *
# 0 1 : F A I # 0 2 : P A S # 0 3 : S T O # 0 4 : - - -
# 0 5 : - - - # 0 6 : - - - # 0 7 : - - - # 0 8 : - - -
# 0 9 : # 1 0 : # 1 1 : # 1 2 :
# 1 3 : # 1 4 : # 1 5 : # 1 6 :
L
-
S
-
P
-
-
-
2. To put the tester back into READY
status, press the STOP button
again.
STOP