User Manual
    GPT-9000/9000A Series User Manual 
80 
DCW Example 
D CW
Primary item
Secondary item
  I ( m A ) :
  V ( k V ) :
  0 . 2 0 0  
  0 0 . 5 1  
  T ( S ) :  
  0 0 0 4 . 5
V
I
Test Time CursorSTA.t time
2.  Use the scroll wheel to move the 
cursor on the time axis (x-axis). 
The measured values for the 
primary and secondary items at 
that particular point in time are 
shown on the left-hand side.  
Remove Lines 
from the Graph 
1.  Pressing the F5 key 
will toggle the 
primary test item 
on/off.  
2.  Pressing the F6 key 
will toggle the 
secondary test item 
on/off. 
D CW
Test Voltage
  I ( m A ) :
  V ( k V ) :
  0 . 2 0 0  
  0 0 . 5 1  
  T ( S ) :  
  0 0 0 4 . 5
V
I
Test Time
Toggle 
voltage line
Toggle 
current line
Example: DCW test 
Exit the Results 
Graph  
To exit the graph, press the ESC 
key. You will be returned back to 
MANU mode/VIEW status. 
ESC