User Manual
  OPERATION 
79 
Running the Test 
1.  In special test mode (000), tests are 
started and stopped in the same 
way as for the normal manual test 
mode. See page 64 for details. 
Page 64 
2.  If required, the scroll wheel can be 
used to set the voltage level in real-
time as the test is running (this 
does not apply to IR or GB tests). 
ACW 
DCW 
0.050kV ~ 5kV 
0.050kV ~ 6kV 
Results 
Test judgments are the same as 
those for the normal manual tests. 
Please see the PASS/FAIL MANU 
Test section for details. 
Page 68 
View Sweep 
Graph 
Unlike normal manual tests, the special test 
mode also has an option to view the resultant 
test as a sweep graph. 
This option is only available for the 
GPT-9901A/9902A/9903/9903A/9904. 
Steps 
1.  When the test has finished, press 
the SWEEP key to view the results 
of the sweep in a graph. 
W E ES P
Graph Test Items: 
TEST 
Primary 
Secondary 
ACW 
Test voltage 
test current 
DCW 
Test voltage 
test current 
IR 
Test voltage 
test resistance 
GB 
Test current 
test resistance 