User Manual
  OPERATION 
77 
Test V
Meas. I
time
Start time 
RAMP TIME
Legend: Voltage:  Current
TEST time
Voltage
Current
The test items that are plotted on the sweep 
graph depend on the type of test that is 
performed. 
TEST 
Graph Test Items 
ACW: 
Test voltage, measured current (V, I) 
DCW: 
Test voltage, measured current (V, I) 
IR: 
Test voltage, measured resistance (V, R) 
GB: 
Test current, measured resistance (I, R) 
Steps 
1.  Choose MANU number 000 to 
enter the special test mode. 
Page 39 
2.  The settings of a previous test can 
be loaded by pressing the 
corresponding soft-key in the 
VIEW or READY status. 
For example, if you are currently in 
DCW mode, pressing the ACW 
key will load the ACW settings 
that were previously used in the 
special manual mode. 
A CW
Example: 
ACW 