Special MANU Test Mode (000) 
Special Test 
Mode Overview 
When MANU number 000 is selected, the 
special test mode is activated. Under the special 
test mode, the voltage can be changed during a 
test, in real time (ACW, DCW only). The test 
function can also be changed when in READY 
or VIEW status, unlike under normal operation. 
Separate settings can be saved under the special 
test mode for each of the testing functions: 
ACW, DCW, IR and GB. This means a different 
ACW, DCW, IR and GB test setup can be saved 
for MANU number 000. 
Sweep Function 
Overview 
The GPT-9901A/9902A/9903/9903A and 9904 
have access to the sweep mode function. The 
sweep function creates a graph of one of the 
ACW, DCW, IR or GB tests in the special 
manual mode. The graph will plot the output 
voltage, current or resistance versus time. After 
the test has been completed, the test current, 
voltage or resistance at any point in time can be 
viewed in the graph. 
Below is an example of the resultant sweep plot 
of a DCW test where a DC voltage is ramped 
up to a user-defined level until the HI SET 
current level has been tripped or the test time 
runs out. 