User Manual
    GPT-9000/9000A Series User Manual 
52 
Setting the ARC Mode 
Background 
ARC detection, otherwise known as flashover 
detection, detects fast voltage or current 
transients that are not normally detected. Arcing 
is usually an indicator of poor withstanding 
insulation, electrode gaps or other insulating 
problems that cause temporary spikes in current 
or voltage during ACW and DCW testing.  
There are three ARC detection settings: OFF, 
ON AND CONTINUE, ON AND STOP. 
The ON AND CONTINUE setting will detect 
arcs over the ARC current level and continue 
the test, the ON AND STOP setting will stop the 
test when an arc is detected. 
ARC mode settings only apply to ACW and 
DCW tests. 
Steps 
1.  Press the UTILITY key on the front 
panel when the tester is in EDIT 
status. The tester will go to the 
MANU Utility for the current test. 
UTILITY
L I T Y    IM A UN U T  2M A N U = * * * - 0 0
 A R C     M D E : OO            FF        
G R O U N D   M O D E : NO
P A S S   H L D :O O FF
F A I L   M D E :O
PS T O
M A X     H O L D : O F F
 Note 
The MANU UTILITY settings only apply to the 
selected MANU test. 
2.  Use the UP/DOWN arrow keys to 
move to the ARC MODE setting. 