Setting the Ramp Up Time 
Background 
The Ramp Up time is the total time taken for 
the tester to reach the test voltage level. The 
Ramp Up time starts after the initial time 
(approximately 150ms) with a start voltage of 
50 volts. The Ramp Up time can be set from 
000.1 to 999.9 seconds. The Ramp Up time is 
only applicable for ACW, DCW and IR tests. 
Discharge time 
Test V
time
Start V
Initial time
TEST TIMERAMP
(Approximately 150ms)
Steps 
1.  Use the UP/DOWN arrow keys to 
bring the cursor to the RAMP  
setting. 
I R   G B    
m A
A CW D C W I / LH O I M ET R
E         
 F R E Q =   0 H z  6
0
100
k  V
IDE
                          MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R  
T
2M A N U = * * * - 0 0
cursor
2.  Use the scroll wheel to set the 
RAMP  value. 
ACW 
DCW 
IR 
000.1s~999.9s 
000.1s~999.9s 
000.1s~999.9s 
 Note 
The discharge time and initial test time cannot be 
edited. 