Setting the Upper and Lower Limits 
Background 
There is both a LO and HI judgment setting. 
When the measured value is below the LO SET 
setting, the test will be judged as FAIL. When 
the value exceeds the HI SET setting the test 
will be judged as FAIL. Any measurement 
between the LO SET and HI SET setting is 
judged as PASS. The LO SET limit cannot be 
made greater than the HI SET limit.  
Steps 
1.  Press the HI/LO soft-key or use 
the UP / DOWN arrow keys to 
bring the cursor to the HI SET 
(ACW/DCW/GB) setting or the 
LO SET(IR) setting. 
I / LH O
OR 
I R   G B    A CW D CW I / LH O I M ET R
E         
 F R E Q =   0 H z  6
0
100
k  V
IDE
                          MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R  
T
2M A N U = * * * - 0 0
cursor
m A
2.  Use the scroll wheel to set the HI 
SET/LO SET limit*.  
ACW (HI) 
DCW (HI) 
IR (LO) 
GB (HI) 
0.001mA~042.0mA (GPT-98XX) 
0.001mA~110.0mA (GPT-99XX/ 
99XXA) 
0.001mA~011.0mA (GPT-98XX) 
0.001mA~021.0mA (GPT-99XX/ 
99XXA) 
0001MΩ ~ 9999MΩ (GPT-98XX) 
0.001GΩ ~ 50.00GΩ (GPT-99XX/ 
99XXA) 
000.1mΩ ~ 650.0mΩ 