User Manual

APPENDIX
173
Table 1b: Output Limitation in Withstanding Voltage Testing (GPT-99XX/99XXA)
Upper Current
Pause
Output Time
AC
80mA≤I≤100mA
At least as long
as the output
time
Maximum 240
seconds
0.001mAI<80
mA
Not necessary
Continuous
output possible
DC
0.001mAI20
mA
Not necessary
Continuous
output possible
GB
15A<I32A
(99XX)
At least as long
as the output
time
999.9
3AI15A
Not necessary
999.9
NOTE: Output Time = Ramp Time + Test Time.