User Manual
  APPENDIX 
171 
Ground Bond Test 
Output Current Range 
03.00A~30.00A (GPT-98XX) 
03.00A~32.00A (GPT-99XX) 
Output Current Accuracy 
 (1% of setting +0.2A) when 3A≤I≤8A 
 (1% of setting +0.05A) when 8A<I≤30A 
(GPT-98XX) 
 (1% of setting +0.05A) when 8A<I≤32A 
(GPT-99XX) 
Output Current Resolution 
0.01A 
Frequency 
50Hz/60Hz selectable 
Ohmmeter Measurement 
Accuracy 
 (1% of reading +2m) 
Ohmmeter Measurement 
Range 
10m~650.0m (depending on output current) 
GPT-98XX: 
GPT-99XX: 
Test Voltage 
Max. 6V(AC)open-circuit 
Ohmmeter Measurement 
Resolution 
0.1m 
Continued… 