User Manual
  APPENDIX 
167 
GPT-9000/9000A Specifications 
The specifications apply when the GPT-9000/9000A is powered on 
for at least 30 minutes at 15˚C~35˚C.  
Specifications 
Environment 
Range 
Temperature 
Humidity 
Warranty 
15˚C ~ 35˚C 
≤70% (No 
condensation) 
Operation 
0˚C ~ 40˚C 
≤70% (No 
condensation) 
Storage 
-10˚C ~ 70˚C 
≤85% (No 
condensation) 
Installation Location 
Indoors at an amplitude of up to 2000m. 
AC Withstanding Voltage 
Output Voltage Range 
0.050kV~ 5.000kV
1
Output Voltage Resolution 
2V 
Output Voltage Accuracy 
 (1% of setting +5V) with no load 
Maximum Rated Load(Table1) 
200 VA (5kV/40mA) [GPT-98XX] 
500 VA (5kV/100mA) [GPT-99XX/99XXA] 
Maximum Rated Current 
40mA [GPT-98XX], 100mA [GPT-99XX/99XXA] 
0.001mA ~ 10mA(0.1kV≤V≤0.5kV) 
0.001mA ~ 40mA(0.5kV<V≤5kV) [GPT-98XX] 
0.001mA ~ 100mA(0.5kV<V≤5kV) [GPT-99XX/ 
99XXA] 
Output Voltage Waveform 
Sine wave 
Frequency 
50 Hz / 60 Hz 
Voltage Regulation 
 1% +5V 
[Maximum rated load → no load] 
Voltmeter Accuracy 
 (1% of reading+ 5V) 
Current Measurement Range 
0.001mA~040.0mA [GPT-98XX] 
0.001mA~100.0mA [GPT-99XX/GPT-99XXA] 
Continued… 