User Manual
  REMOTE CONTROL 
145 
Syntax 
Query Syntax 
MANU:GB:TTIMe <NR2> 
MANU:GB:TTIMe? 
Parameter/ 
Return parameter 
<NR2> 
0.5 ~ 999.9 seconds 
Example 
MANU:GB:TTIM 1 
Sets the GB test time to 1 second. 
MANU:GB:FREQuency 
Set
Query
Description 
Sets or returns the GB test frequency in Hz. The 
test must first be in GB mode before this command 
can be used. 
Syntax 
Query Syntax 
MANU:GB:FREQuency {50|60} 
MANU:GB:FREQuency? 
Parameter/ 
Return parameter 
50 
50 Hz 
60 
60 Hz 
Example 
MANU:GB:FREQ 50 
Sets the GB test frequency to 50Hz. 
MANU:GB:REF 
Set
Query
Description 
Sets or returns the GB reference value in mΩ. The 
test must first be in GB mode before this command 
can be used. 
Syntax 
Query Syntax 
MANU:GB:REF <NR2> 
MANU:GB:REF? 
Parameter/ 
Return parameter 
<NR2> 
0.000 ~ 649.9 
Example 
MANU:GB:REF 100 
Sets the GB reference to 100 mΩ. 