User Manual
REMOTE CONTROL
145
Syntax
Query Syntax
MANU:GB:TTIMe <NR2>
MANU:GB:TTIMe?
Parameter/
Return parameter
<NR2>
0.5 ~ 999.9 seconds
Example
MANU:GB:TTIM 1
Sets the GB test time to 1 second.
MANU:GB:FREQuency
Set
Query
Description
Sets or returns the GB test frequency in Hz. The
test must first be in GB mode before this command
can be used.
Syntax
Query Syntax
MANU:GB:FREQuency {50|60}
MANU:GB:FREQuency?
Parameter/
Return parameter
50
50 Hz
60
60 Hz
Example
MANU:GB:FREQ 50
Sets the GB test frequency to 50Hz.
MANU:GB:REF
Set
Query
Description
Sets or returns the GB reference value in mΩ. The
test must first be in GB mode before this command
can be used.
Syntax
Query Syntax
MANU:GB:REF <NR2>
MANU:GB:REF?
Parameter/
Return parameter
<NR2>
0.000 ~ 649.9
Example
MANU:GB:REF 100
Sets the GB reference to 100 mΩ.