MANU:GB:RHISet 
Set
Query
Description 
Sets or returns the GB HI SET resistance value in 
mΩ. The test must first be in GB mode before this 
command can be used. 
Syntax 
Query Syntax 
MANU:GB:RHISet <NR2> 
MANU:GB:RHISet? 
Parameter/ 
Return parameter 
<NR2> 
000.1 ~ 650.0 
Example 
MANU:GB:RHIS 100.0 
Sets the HI SET value to 100mΩ. 
 Note 
If the (GB current x HI SET resistance) > 5.4V, then an 
error will be generated (“GBV > 5.4V”). 
MANU:GB:RLOSet 
Set
Query
Description 
Sets or returns the GB LO SET resistance value in 
mΩ. The LO SET value must be less than the HI 
SET value. The test must first be in GB mode 
before this command can be used. 
Syntax 
Query Syntax 
MANU:GB:RLOSet<NR2> 
MANU:IR:RLOSet? 
Parameter/ 
Return parameter 
<NR2> 
0.000 ~ 649.9 
Example 
MANU:GB:RLOS 50 
Sets the GB LO SET resistance to 50mΩ. 
MANU:GB:TTIMe 
Set
Query
Description 
Sets or returns the GB test time in seconds. The test 
must first be in GB mode before this command can 
be used. 