User Manual

GPT-9000/9000A Series User Manual
142
Example
(GPT-99XX/
99XXA)
MANU:IR:RHIS 0.010
Sets the IR HI SET resistance to 10 M.
MANU:IR:RLOSet
Set
Query
Description
Sets or returns the IR LO SET resistance value in
M (GPT-98XX) or GΩ. The LO SET value must be
less than the HI SET value. The test must first be
in IR mode before this command can be used.
Syntax
Query Syntax
MANU:IR:RLOSet<NR1>
MANU:IR:RLOSet?
Parameter/
Return parameter
<NR1>
GPT-98XX only:
1 ~ 9999 (unit = M)
GPT-99XX/GPT-99XXA only:
Format A: 0.001 ~ 50.00 (unit = G)
Format B: 0.001G ~ 50.00G
Format C: 1M ~ 50000M
Example
(GPT-98XX)
MANU:IR:RLOS 10
Sets the IR LO SET resistance to 10M.
Example
(GPT-99XX/
99XXA)
MANU:IR:RLOS 0.010
Sets the IR LO SET resistance to 10M.
MANU:IR:TTIMe
Set
Query
Description
Sets or returns the IR test time in seconds. The test
must first be in IR mode before this command can
be used.
Syntax
Query Syntax
MANU:IR:TTIMe <NR2>
MANU:IR:TTIMe?
Parameter/
Return parameter
<NR2>
1.0 ~ 999.9 seconds
Example
MANU:IR:TTIM 1
Sets the IR test time to 1 second.