User Manual
    GPT-9000/9000A Series User Manual 
140 
Example 
MANU:DCW:TTIM 1 
Sets the DCW test time to 1 second. 
MANU:DCW:REF 
Set
Query
Description 
Sets or returns the DCW reference value in mA. 
The test must first be in DCW mode before this 
command can be used. 
The reference value must be less than the HI SET 
value. 
The reference value uses the same range as the HI 
SET value. 
Syntax 
Query Syntax 
MANU:DCW:REF <NR2> 
MANU:DCW:REF? 
Parameter/ 
Return parameter 
<NR2> 
0.000 ~ 010.9 (GPT-98XX) 
0.000 ~ 020.9 (GPT-99XX/99XXA) 
Example 
MANU:DCW:REF 0.01 
Sets the DCW reference to 0.01 mA. 
MANU:DCW:ARCCurrent 
Set
Query
Description 
Sets or returns the DCW ARC current value in mA. 
ARC must be enabled to set the ARC current. The 
test must first be in DCW mode before this 
command can be used. 
ARC current uses the same range as the HI SET 
value. The ARC current is limited to 2X the HI SET 
value. 
Syntax 
Query Syntax 
MANU:DCW:ARCCurrent <NR2> 
MANU:DCW:ARCCurrent? 
Parameter/ 
Return parameter 
<NR2> 
1.000 ~ 20.00 (GPT-98XX) 
2.000 ~ 040.0 (GPT-99XX/99XXA) 