User Manual
REMOTE CONTROL
139
an error as all digits are outside the range of HI
SET. LO SET value2 will not produce an error, but
will return 0.05, not 0.053.
Syntax
Query Syntax
MANU:DCW:CLOSet<NR2>
MANU:DCW:CLOSet?
Parameter/
Return parameter
<NR2>
0.000 ~ 010.9 (GPT-98XX)
0.000 ~ 020.9 (GPT-99XX/99XXA)
Example
MANU:DCW:CLOS 2.00
Sets the DCW LO SET current to 2mA.
MANU:DCW:TTIMe
Set
Query
Description
Sets or returns the DCW test time in seconds. The
test must first be in DCW mode before this
command can be used.
In special MANU mode, the TIMER can be turned
off.
Syntax
Query Syntax
MANU:DCW:TTIMe {<NR2>|OFF}
MANU:DCW:TTIMe?
Parameter
<NR2>
OFF
0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).
Return parameter
<NR2>
TIME OFF
0.5 ~ 999.9 seconds
TIMER is OFF (special MANU mode).