  REMOTE CONTROL 
139 
an error as all digits are outside the range of HI 
SET. LO SET value2 will not produce an error, but 
will return 0.05, not 0.053. 
Syntax 
Query Syntax 
MANU:DCW:CLOSet<NR2> 
MANU:DCW:CLOSet? 
Parameter/ 
Return parameter 
<NR2> 
0.000 ~ 010.9 (GPT-98XX) 
0.000 ~ 020.9 (GPT-99XX/99XXA) 
Example 
MANU:DCW:CLOS 2.00 
Sets the DCW LO SET current to 2mA. 
MANU:DCW:TTIMe 
Set
Query
Description 
Sets or returns the DCW test time in seconds. The 
test must first be in DCW mode before this 
command can be used. 
In special MANU mode, the TIMER can be turned 
off. 
Syntax 
Query Syntax 
MANU:DCW:TTIMe {<NR2>|OFF} 
MANU:DCW:TTIMe? 
Parameter 
<NR2> 
OFF 
0.5 ~ 999.9 seconds 
TIMER OFF (special MANU mode). 
Return parameter 
<NR2> 
TIME OFF 
0.5 ~ 999.9 seconds 
TIMER is OFF (special MANU mode). 