138 
MANU:DCW:CHISet 
Set
Query
Description 
Sets or returns the DCW HI SET current value in 
milliamps. The test must first be in DCW mode 
before this command can be used. 
Note: A “DC Over 50W” error will result if the 
DCW Voltage X HI SET value is > 50 watts. 
Note: A “DC Over 100W” error will result if the 
DCW Voltage X HI SET value is > 100 watts (GPT-
99XX/99XXA) 
Syntax 
Query Syntax 
MANU:DCW:CHISet <NR2> 
MANU:DCW:CHISet? 
Parameter/ 
Return parameter 
<NR2> 
0.001 ~ 011.0 (GPT-98XX) 
0.001 ~ 021.0 (GPT-99XX/99XXA) 
Example 
MANU:DCW:CHIS 5 
Sets the DCW HI SET current to 5mA. 
MANU:DCW:CLOSet 
Set
Query
Description 
Sets or returns the DCW LO SET current value in 
milliamps. The LO SET value must be less than the 
HI SET value. The test must first be in DCW mode 
before this command can be used. 
The LO SET range must use the HI SET range. If all 
the digits in the LO SET range are outside the HI 
SET range, an error will be produced. All digits 
outside the HI SET range are ignored and will not 
be used. 
For example: 
HI SET value: 12.34 
LO SET value1: 0.005  error 
LO SET value2: 0.053  no error 
In the example above LO SET value1 will produce 