User Manual
  REMOTE CONTROL 
137 
MANU:ACW:ARCCurrent 
Set
Query
Description 
Sets or returns the ACW ARC current value in mA. 
ARC must be enabled before the ARC current can 
be set. The test must first be in ACW mode before 
this command can be used. 
ARC current uses the same range as the HI SET 
value. The ARC current is limited to 2X the HI SET 
value. 
Syntax 
Query Syntax 
MANU:ACW:ARCCurrent <NR2> 
MANU:ACW:ARCCurrent? 
Parameter/ 
Return parameter 
<NR2> 
1.000 ~ 080.0 (GPT-98XX) 
2.000 ~ 200.0 (GPT-99XX/99XXA) 
Example 
MANU:ACW:ARCC 0.04 
Sets the ACW ARC value to 0.04 mA. 
MANU:DCW:VOLTage 
Set
Query
Description 
Sets or returns the DCW voltage in kV. The test 
must first be in DCW mode before this command 
can be used. 
Note: A “DC Over 50W” error will result if the 
DCW Voltage X HI SET value is > 50 watts (GPT-
98XX). 
Note: A “DC Over 100W” error will result if the 
DCW Voltage X HI SET value is > 100 watts (GPT-
99XX/99XXA).   
Syntax 
Query Syntax 
MANU:DCW:VOLTage <NR2> 
MANU:DCW:VOLTage? 
Parameter/ 
Return parameter 
<NR2> 
0.050 ~ 6.100 (kV) 
Example 
MANU:DCW:VOLT 6 
Sets the DCW voltage to 6 kV. 