User Manual
  REMOTE CONTROL 
135 
Syntax 
Query Syntax 
MANU:ACW:CLOSet<NR2> 
MANU:ACW:CLOSet? 
Parameter/ 
Return parameter 
<NR2> 
0.000 ~ 041.9 (GPT-98XX) 
0.000 ~ 109.9 (GPT-99XX/99XXA) 
Example 
MANU:ACW:CLOS 20.0 
Sets the ACW LO SET current to 20 mA. 
MANU:ACW:TTIMe 
Set
Query
Description 
Sets or returns the ACW test time in seconds. The 
test must first be in ACW mode before this 
command can be used. 
Note: A “TIME ERR” will result if the Ramp Time 
+ Test Time is ≥ 240 seconds when the HI SET limit 
is over 30mA (GPT-98XX) or over 80mA (GPT-
99XX/99XXA). This applies to the ACW function 
only. 
In special MANU mode, the TIMER can be turned 
off. 
Syntax 
Query Syntax 
MANU:ACW:TTIMe {<NR2>|OFF} 
MANU:ACW:TTIMe? 
Parameter 
<NR2> 
OFF 
0.5 ~ 999.9 seconds 
TIMER OFF (special MANU mode). 
Return parameter 
<NR2> 
TIME OFF 
0.5 ~ 999.9 seconds 
TIMER is OFF (special MANU 
mode). 
Example 
MANU:ACW:TTIM 1 
Sets the ACW test time to 1 second. 