User Manual
  REMOTE CONTROL 
133 
Parameter/ 
Return parameter 
ACW 
AC Withstand mode 
DCW 
DC Withstand mode 
IR 
Insulation Resistance mode 
GB 
Ground Bond mode 
Example 
MANU:EDIT:MODE ACW 
Sets the mode to ACW. 
MANU:ACW:VOLTage 
Set
Query
Description 
Sets or returns the ACW voltage in kV. The test 
must first be in ACW mode before this command 
can be used. 
Syntax 
Query Syntax 
MANU:ACW:VOLTage <NR2> 
MANU:ACW:VOLTage? 
Parameter/ 
Return parameter 
<NR2> 
0.050 ~ 5.000 (kV) 
Example 
MANU:ACW:VOLT 1 
Sets the ACW voltage to 1 kV. 