User Manual
REMOTE CONTROL
133
Parameter/
Return parameter
ACW
AC Withstand mode
DCW
DC Withstand mode
IR
Insulation Resistance mode
GB
Ground Bond mode
Example
MANU:EDIT:MODE ACW
Sets the mode to ACW.
MANU:ACW:VOLTage
Set
Query
Description
Sets or returns the ACW voltage in kV. The test
must first be in ACW mode before this command
can be used.
Syntax
Query Syntax
MANU:ACW:VOLTage <NR2>
MANU:ACW:VOLTage?
Parameter/
Return parameter
<NR2>
0.050 ~ 5.000 (kV)
Example
MANU:ACW:VOLT 1
Sets the ACW voltage to 1 kV.