User Manual
    GPT-9000/9000A Series User Manual 
132 
Initial Settings 
Parameter 
Function 
ACW 
DCW 
IR 
GB 
REF# 
0.000mA 
0.000mA 
0000MΩ 
000.0mΩ 
FREQ 
60Hz 
X 
X 
60Hz 
HI SET 
1.000mA 
1.000mA 
∞MΩ 
100.0mΩ 
LO SET 
0.000mA 
0.000mA 
0001MΩ 
000.0mΩ 
I or V 
V=0.100kV 
V=0.100kV 
V=0.050kV 
03.00A 
TIMER 
001.0S 
001.0S 
001.0S 
001.0S 
RAMP↗ 
000.1S 
000.1S 
000.1S 
X 
Example 
MANU:INITial 
Loads the initial settings for the selected MANU 
number. 
MANU:RTIMe 
Set
Query
Description 
Sets or returns the Ramp Time for the test in 
seconds. 
Note: A “TIME ERR” will result if the Ramp Time 
+ Test Time is ≥ 240 seconds when the HI SET limit 
is over 30mA (GPT-98XX) or over 80mA (GPT-
99XX/99XXA). This applies to the ACW function 
only.  
Syntax 
Query Syntax 
MANU:RTIMe <NR2> 
MANU:RTIMe? 
Parameter/ 
Return parameter 
<NR2> 
0.1~999.9 seconds 
Example 
MANU:RTIM 0.5 
Sets the ramp time to half a second. 
MANU:EDIT:MODE 
Set
Query
Description 
Sets or returns the mode (ACW, DCW, IR, GB) of 
the selected manual test. 
Syntax 
Query Syntax 
MANU:EDIT:MODE {ACW|DCW|IR|GB} 
MANU:EDIT:MODE? 