User Manual
GPT-9000/9000A Series User Manual
132
Initial Settings
Parameter
Function
ACW
DCW
IR
GB
REF#
0.000mA
0.000mA
0000MΩ
000.0mΩ
FREQ
60Hz
X
X
60Hz
HI SET
1.000mA
1.000mA
∞MΩ
100.0mΩ
LO SET
0.000mA
0.000mA
0001MΩ
000.0mΩ
I or V
V=0.100kV
V=0.100kV
V=0.050kV
03.00A
TIMER
001.0S
001.0S
001.0S
001.0S
RAMP↗
000.1S
000.1S
000.1S
X
Example
MANU:INITial
Loads the initial settings for the selected MANU
number.
MANU:RTIMe
Set
Query
Description
Sets or returns the Ramp Time for the test in
seconds.
Note: A “TIME ERR” will result if the Ramp Time
+ Test Time is ≥ 240 seconds when the HI SET limit
is over 30mA (GPT-98XX) or over 80mA (GPT-
99XX/99XXA). This applies to the ACW function
only.
Syntax
Query Syntax
MANU:RTIMe <NR2>
MANU:RTIMe?
Parameter/
Return parameter
<NR2>
0.1~999.9 seconds
Example
MANU:RTIM 0.5
Sets the ramp time to half a second.
MANU:EDIT:MODE
Set
Query
Description
Sets or returns the mode (ACW, DCW, IR, GB) of
the selected manual test.
Syntax
Query Syntax
MANU:EDIT:MODE {ACW|DCW|IR|GB}
MANU:EDIT:MODE?