User Manual
  REMOTE CONTROL 
125 
SYSTem:BUZZer:FTIMe 
Set
Query
Description 
Sets the FAIL Sound duration in seconds. 
Syntax 
Query Syntax 
SYSTem:BUZZer:FTIMe <NR2> 
SYSTem:BUZZer:FTIMe? 
Parameter/ 
Return parameter 
<NR2> 
0.2~999.9 
Example 
SYST:BUZZ:FTIM 1 
Sets the buzzer to 1 second for a FAIL judgment. 
SYSTem:ERRor 
Query
Description 
Returns any errors in the output buffer. See the 
error code table below for details. 
Query Syntax 
SYSTem:ERRor ? 
Return parameter 
<string> 
Returns an error string that includes 
an error code and an error 
description. 
Error Code Table 
Error code, Error description 
0,No Error 
20,Command Error 
21,Volume Error 
22,String Error 
23,Query Error 
24,Mode Error 
25,Time Error 
26,DC Over 50W (GPT-98XX), DC Over 100W (GPT- 
 99XX/99XXA 
27,GBV > 5.4V 
30,Voltage Setting Error 
31,Current Setting Error 
32,Current HI SET Error 
33,Current LOW SET Error 
34,Resistance HI SET Error 