User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
98
Remote Controller Operation
Description
The GPT-9000 accepts external remote
controllers with a START and STOP button. To
use the REMOTE terminal, the GPT-9000 must
first be configured to accept a remote controller.
Operating a remote controller is the same as
operating the START and STOP buttons on the
front panel.
Steps
1. Turn off the power.
2. Insert the lead of remote controller
into the REMOTE terminal.
REMOTE
3. Turn on the power to the tester.
4. Configure the Start Ctrl option to
REMOTE CONNECT in the
Common Utility menu.
Page 76
5. The tester will now only be able to
start a test using a remote
controller.
6. To return the operation control to
the front panel, configure the Start
Ctrl option to FRONT PANEL.
Page 76