User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
70
Special MANU Test Mode (000)
Background
When MANU number 000 is selected, the
special test mode is activated. Under the special
test mode, the voltage can be changed during a
test, in real time (ACW, DCW only). The test
function can also be changed when in READY
or VIEW status, unlike under normal operation.
Separate settings can be saved under the special
test mode for each of the testing functions:
ACW, DCW, IR and GB. This means a different
ACW, DCW, IR and GB test setup can be saved
for MANU number 000.
Steps
1. Choose MANU number 000 to
enter the special test mode.
Page 36
2. Set all the necessary parameters for
a test and save. A different test
setup can be saved for each test
function (ACW, DCW, IR and GB).
Pages 37~57
I R G B A CW D C W
E
F R E Q = 0 H z 6
0
100
k V
RA C = F FO
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
MM A UN N A_0M A N U = * * * - 0 0
m A
Note
The TIMER settings can be set OFF when in the
special test mode for ACW and DCW tests.
3. The settings of a different test
function can be loaded by pressing
the corresponding soft-key in the
VIEW or READY status.
A CW
Example:
ACW