User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
48
3. Use the scroll wheel to set the ARC
mode.
ARC MODES:
OFF, ON AND CONTINUE,
ON AND STOP
4. Press the EDIT/SAVE key to save
and exit the MANU Utility and go
back to EDIT status.
EDIT/SAVE
Note
The ESC key can be pressed at any time in the
Utility menu to cancel and exit.
5. If the ARC MODE was set to either ON AND
CONTINUE, or ON AND STOP, the ARC
current level can be edited.
6. Use the UP/DOWN arrow keys to
move the cursor to the ARC setting.
7. Use the scroll wheel to edit the
ARC level.
ACW
DCW
1.000mA~080.0mA
1.000mA~020.0mA
Note
The ARC setting range is directly related to the HI
SET current limit.
ACW
HI SET Limit
ARC range
0.001mA~0.999mA
1.000mA ~2.000mA
01.00mA~09.99mA
01.00mA ~20.00mA
010.0mA~042.0mA
001.0mA ~080.0mA