User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
OPERATION
41
I R G B A C W D CW I / LH O I M ET R
E
F R E Q = 0 H z 6
0
100
k V
RA C = F FO
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
m A
2. Use the scroll wheel to set the HI
SET/LO SET limit.
ACW (HI)
DCW (HI)
IR (LO)
GB (HI)
0.001mA~042.0mA
0.001mA~011.0mA
0001MΩ ~ 9999MΩ
000.1mΩ ~ 650.0mΩ
3. Press the HI/LO soft-key again or
press the DOWN arrow key to
switch between HI SET and LO
SET.
I / LH O
OR
I R G B A CW D C W I / LH O I M ET R
E
F R E Q = 0 H z 6
0
100
k V
RA C = F FO
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
1 . 0 0 m AIH ES = 0T
m A
4. Use the scroll wheel to set the HI
SET/LO SET limit.
ACW (LO)
DCW (LO)
IR (HI)
GB (LO)
0.000mA~041.9mA
0.000mA~010.9mA
0001MΩ~9500MΩ, ∞
000.0mΩ ~ 649.9mΩ
Note
The LO SET setting is limited by the HI SET
setting. The LO SET limit cannot be greater than
the HI SET limit.