User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
APPENDIX
151
RAMP (Ramp Time)
0.1~999.9S
TIMER (Test Time)
1S~999.9S
GND
OFF
Ground Bond Test
Output Current Range
03.00A~30.00A
Output Current Accuracy
(1% of reading +0.2A) when 3A≤I≤8A
(1% of reading +0.05A) when 8A<I≤30A
Output Current Resolution
0.01A
Frequency
50Hz/60Hz
Ohmmeter Measurement
Accuracy
(1% of reading +2m)
Ohmmeter Measurement
Range
10m~650.0m
Test Voltage
Max. 6V(AC)
Ohmmeter Measurement
Resolution
0.1m
Windows Comparator Method
Yes
TIMER (Test Time)
0.5S~999.9S
GND
OFF
Interface
REMOTE (Remote terminal)
Yes
SIGNAL IO
Yes
RS232
Yes
USB (Device)
Yes
GPIB
Yes (OPTION)
General
DISPLAY
240 x 64 dot matrix LED back light LCD
MEMORY
AUTO/MANU mode 100 memory blocks total
Continued…