User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
150
TIMER (Test Time)
OFF*, 0.5S~999.9S
GND
ON/OFF
* The timer can only be turned off under special MANU mode (MANU=***-000)
DC Withstanding Voltage
Output Voltage Range
0.100kV~ 6.000kV
Output Voltage Resolution
2V
Output Voltage Accuracy
(1% of setting +5V) with no load
Maximum Rated Load(Table1)
50W (5kV/10mA)
Maximum Rated Current
10mA
0.001mA ~ 2mA (0.1kV≤V≤0.5kV)
0.001mA ~ 10mA (0.5kV<V≤6kV)
Voltmeter Accuracy
(1% of reading+ 5V)
Voltage Regulation
1% +5V
[Maximum rated load → no load]
Current Measurement Range
0.001mA~010.0mA
Current Best Resolution
1uA
0.001mA(0.001mA~0.999mA)
0.01mA(01.00mA~09.99mA)
0.1mA(010.0mA)
Current Measurement
Accuracy
± (1.5% of reading +30uA)
Window Comparator Method
Yes
ARC DETECT
Yes
Rise-time Control Function
Yes
RAMP (Ramp Time)
0.1~999.9S
TIMER (Test Time)
OFF*, 0.5S~999.9S
GND
ON/OFF
* The timer can only be turned off under special MANU mode (MANU=***-000)
Insulation Resistance Test
Output Voltage
50V~1000V
Output Voltage Resolution
50V
Output Voltage Accuracy
(1% of setting+5V) with no load
Resistance Measurement
Range
1M~ 9500M
Test Voltage
Measurement Range
Accuracy
50V≤V<500V
1~50MΩ
±(5% of reading +1MΩ)
51~2000MΩ
±(10% of reading +1MΩ)
500V≤V≤1000V
1~500MΩ
±(5% of reading +1MΩ)
501~9500MΩ
±(10% of reading +1MΩ)
Voltage Regulation
±1% +5V [Maximum rated load → no load]
Window Comparator Method
Yes
Rise-time Control Function
Yes
Continued…