User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
APPENDIX
149
GPT-9000 Specifications
The specifications apply when the GPT-9000 is powered on for at
least 30 minutes at 15˚C~35˚C.
Specifications
Environment
Range
Temperature
Humidity
Warranty
15˚C ~ 35˚C
≤70% (No
condensation)
Operation
0˚C ~ 40˚C
≤70% (No
condensation)
Storage
-10˚C ~ 70˚C
≤85% (No
condensation)
Installation Location
Indoors at an amplitude of up to 2000m.
AC Withstanding Voltage
Output Voltage Range
0.100kV~ 5.000kV
Output Voltage Resolution
2V
Output Voltage Accuracy
(1% of setting +5V) with no load
Maximum Rated Load(Table1)
200 VA (5kV/40mA)
Maximum Rated Current
40mA
0.001mA ~ 10mA(0.1kV≤V≤0.5kV)
0.001mA ~ 40mA(0.5kV<V≤5kV)
Output Voltage Waveform
Sine wave
Frequency
50 Hz / 60 Hz
Voltage Regulation
1% +5V
[Maximum rated load → no load]
Voltmeter Accuracy
(1% of reading+ 5V)
Current Measurement Range
0.001mA~040.0mA
Current Best Resolution
1uA
0.001mA(0.001mA~0.999mA)
0.01mA(01.00mA~09.99mA)
0.1mA(010.0~040.0mA)
Current Measurement
Accuracy
± (1.5% of rdg + 30 counts) when HI SET<1.00mA
± (1.5% of rdg + 3 counts) when HI SET≥1.00mA
Window Comparator Method
Yes
ARC DETECT
Yes
Rise-time Control Function
Yes
RAMP (Ramp Time)
0.1~999.9S
Continued…