User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
REMOTE CONTROL
135
Parameter/
Return parameter
OFF
Turns FAIL HOLD off.
ON
Turns FAIL HOLD on.
Example
MANU:UTIL:FAIL OFF
Turns FAIL HOLD OFF.
MANU:UTILity:MAXHold
Set
Query
Description
Sets or returns the MAX HOLD setting for the
current test.
Syntax
Query Syntax
MANU:UTILity:MAXHold {ON|OFF}
MANU:UTILity:MAXHold?
Parameter/
Return parameter
OFF
Turns MAX HOLD off.
ON
Turns MAX HOLD on.
Example
MANU:UTIL:MAXH ON
Turns MAX HOLD on.
MANU:UTILity:GROUNDMODE
Set
Query
Description
Sets or returns the Grounding mode of the current
test.
The Ground Mode setting cannot be turned on
with the IR and GB function.
Syntax
Query Syntax
MANU:UTILity:GROUNDMODE {ON|OFF}
MANU:UTILity:GROUNDMODE?
Parameter/
Return parameter
OFF
Turns ground mode off.
ON
Turns ground mode on.
Example
MANU:UTIL:GROUNDMODE ON
Turns GROUND MODE on.