User manual

Table Of Contents
GPT-9000 Series User Manual
134
MANU:UTILity:ARCMode
Set
Query
Description
Sets or returns the ARC mode status for the
current test.
The ARC mode cannot be set for the IR and GB
function.
Syntax
Query Syntax
MANU:UTILity:ARCMode {OFF|ON_CONT|
ON_STOP}
MANU:UTILity:ARCMode?
Parameter/
Return parameter
OFF
Turns ARC mode off.
ON_CONT
Sets ARC mode to ON and
CONTINUE.
ON_STOP
Sets ARC mode to ON and STOP.
Example
MANU:UTIL:ARCM OFF
Turns ARC mode OFF.
MANU:UTILity:PASShold
Set
Query
Description
Sets or returns the PASS HOLD setting for the
current test.
Syntax
Query Syntax
MANU:UTILity:PASShold {ON|OFF}
MANU:UTILity:PASShold?
Parameter/
Return parameter
OFF
Turns PASS HOLD off.
ON
Turns PASS HOLD on.
Example
MANU:UTIL:PASS OFF
Turns PASS HOLD OFF.
MANU:UTILity:FAILhold
Set
Query
Description
Sets or returns the FAIL HOLD setting for the
current test.
Syntax
Query Syntax
MANU:UTILity:FAILhold {ON|OFF}
MANU:UTILity:FAILhold?