User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
130
MANU:IR:TTIMe
Set
Query
Description
Sets or returns the IR test time in seconds. The test
must first be in IR mode before this command can
be used.
Syntax
Query Syntax
MANU:IR:TTIMe <NR2>
MANU:IR:TTIMe?
Parameter/
Return parameter
<NR2>
1.0 ~ 999.9 seconds
Example
MANU:IR:TTIM 1
Sets the IR test time to 1 second.
MANU:IR:REF
Set
Query
Description
Sets or returns the IR reference value in MΩ. The
test must first be in IR mode before this command
can be used.
The reference value must be lower than the HI SET
value.
Syntax
Query Syntax
MANU:IR:REF <NR1>
MANU:IR:REF?
Parameter/
Return parameter
<NR1>
0000 ~ 9999
Example
MANU:IR:REF 900
Sets the IR reference to 900 MΩ.
MANU:GB:CURRent
Set
Query
Description
Sets or returns the GB current in A. The test must
first be in GB mode before this command can be
used.