GPT-9000 Series User Manual
130
MANU:IR:TTIMe
Set
Query
Description
Sets or returns the IR test time in seconds. The test
must first be in IR mode before this command can
be used.
Syntax
Query Syntax
MANU:IR:TTIMe <NR2>
MANU:IR:TTIMe?
Parameter/
Return parameter
<NR2>
1.0 ~ 999.9 seconds
Example
MANU:IR:TTIM 1
Sets the IR test time to 1 second.
MANU:IR:REF
Set
Query
Description
Sets or returns the IR reference value in M. The
test must first be in IR mode before this command
can be used.
The reference value must be lower than the HI SET
value.
Syntax
Query Syntax
MANU:IR:REF <NR1>
MANU:IR:REF?
Parameter/
Return parameter
<NR1>
0000 ~ 9999
Example
MANU:IR:REF 900
Sets the IR reference to 900 M.