User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
REMOTE CONTROL
129
Example
MANU:IR:VOLT 1
Sets the IR voltage to 1 kV.
MANU:IR:RHISet
Set
Query
Description
Sets or returns the IR HI SET resistance value in
MΩ. The test must first be in IR mode before this
command can be used.
Syntax
Query Syntax
MANU:IR:RHISet <NR1>|NULL
MANU:IR:RHISet?
Parameter/
Return parameter
<NR1>
2 ~ 9999
NULL
Sets the HI SET value to high impedance
Example
MANU:IR:RHIS 10.
Sets the IR HI SET resistance to 10 MΩ.
MANU:IR:RLOSet
Set
Query
Description
Sets or returns the IR LO SET resistance value in
MΩ. The LO SET value must be less than the HI
SET value. The test must first be in IR mode before
this command can be used.
Syntax
Query Syntax
MANU:IR:RLOSet<NR1>
MANU:IR:RLOSet?
Parameter/
Return parameter
<NR1>
1 ~ 9999
Example
MANU:IR:RLOS 10
Sets the IR LO SET resistance to 10MΩ.