User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
REMOTE CONTROL
117
MEASure<x>
Query
Description
Returns the test parameters & results of the tester
in either MANU or AUTO mode.
MANU mode: Returns the test parameters &
results of a MANU test.
AUTO mode: Returns the test parameters & results
of the selected step (1-16) of the AUTO test.
Return parameters: function, judgment/status, test
voltage, test current/resistance, test time (time of
completed test) or ramp time (elapsed time of test
that has not been completed.
Query Syntax
MEASure<x>?
Parameter
(MANU mode)
No parameter needed for MANU
mode.
Parameter
(AUTO mode)
<x>
<NR1>1~16. Step number.
Return parameter
<string>
Returns the test status of the test
in the following format:
function, judgment or status, test
voltage, test current or resistance,
test time or ramp time
Function
ACW, DCW, IR, GB
Judgment
/Status
PASS, FAIL
VIEW
Test voltage
voltage+unit
Test current
/Test resistance
current+unit
resistance+unit
Test time
/Ramp time
T=time+S
R=time+S
Example
(in MANU mode)
MEAS?
>ACW, FAIL , 0.024kV ,0.013 mA ,R=000.1S
Returns the test result of the current manual test.