User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL HOLD
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Common Utility Settings
- Automatic Tests
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
REMOTE CONTROL
115
SYSTem:BUZZer:FTIMe
Set
Query
Description
Sets the FAIL Sound duration in seconds.
Syntax
Query Syntax
SYSTem:BUZZer:FTIMe <NR2>
SYSTem:BUZZer:FTIMe?
Parameter/
Return parameter
<NR2>
0.2~999.9
Example
SYST:BUZZ:FTIM 1
Sets the buzzer to 1 second for a FAIL judgment.
SYSTem:ERRor
Query
Description
Returns any errors in the output buffer. See page
143 for a list of all the possible errors.
Query Syntax
SYSTem:ERRor ?
Return parameter
<string>
Returns an error string.
Example
SYST:ERR ?
>Value Error!
Returns “Value Error!” as the error message.
SYSTem:GPIB:VERSion
Query
Description
Queries the GPIB version.
Query Syntax
SYSTem:GPIB:VERSion?
Return parameter
<string>
Returns:
The GPIB version as a string
“GPIB,V1.00”
or
“No GPIB connected” if there is not a
GPIB device configured/connected.
Query Example
SYST:GPIB:VERS?
>GPIB,V1.00
Returns the GPIB version.