APPENDIX
71
Output Limitation for
Withstanding Voltage Testing
Upper Current
Pause
Output Time
15mA≤I≤20mA
At least as long
as the output
time
Maximum 180
seconds
0.01mA≤I<15mA
Not necessary
Continuous
output
possible
DC Withstanding Voltage
Output Voltage Range
0.10kV~ 6.00kV
Output Voltage Resolution
10V
Maximum Rated Load
25W (5kV/5mA)
Maximum Rated Current
6mA
0.01mA ~ 2mA (0.1kV≤V≤0.5kV)
0.01mA ~ 6mA (0.5kV<V≤6kV)
Voltmeter Accuracy
±(1.5% of reading + 2 counts)
Output Voltage Accuracy
±(1.5% of setting + 2 counts) with no load
Voltage Regulation
± 1.5% +2 counts
[Maximum rated load → no load]
Current Measurement Range
0.01mA~6.00mA
Current Best Resolution
10uA
0.01mA(0.01mA~6.00mA)
Current Measurement
Accuracy
± 2% of reading + 5 counts when I <1.00mA
± 2% of reading + 3 counts when I ≥1.00mA
Current Judgment Accuracy
± 3% of setting + 5 counts when I <1.00mA
± 3% of setting + 3 counts when I ≥1.00mA
Window Comparator Method
Yes
ARC DETECT
Yes
RAMP (Ramp Time)
0.1S fixed
TIMER (Test Time)
OFF, 1S~180S
GND
ON
Judgment Valid Range
H/L Setting: 0.10mA to 6.00mA
Insulation Resistance Test
Output Voltage
50V, 100V, 250V, 500V, 1000V
Output Voltage Accuracy
±(3% of setting+1 count) with no load
Resistance Measurement
Range
1M ~ 2000M
Test Voltage
Measurement Range
Accuracy
50V/100V/250
1 ~ 50MΩ
±(5% of reading + 2MΩ)
51~2000MΩ
±(10% of reading + 2MΩ)
500V/1000V
1 ~ 500MΩ
±(5% of reading + 2MΩ)
501~2000MΩ
±(10% of reading + 2MΩ)
Voltmeter accuracy
±(3% of reading + 1 count)
Continued…