GPT-9600 Series User Manual
36
ACW (H)
DCW (H)
IR (H)
0.01mA~22.0mA
0.01mA~6.00mA
0002MΩ~∞Ω
3. Press the FIELD soft-key to bring
the cursor to the L setting or the H
setting.
FIELD
H = 2 2 . 0 m A
m AL 0= 0 0. 7 4 S = 1T
CWDA CW R I DO EM T M EI
cursor
1 00
k V
IDE
T
Am
4. Use the scroll wheel to set the H/L
limit*.
ACW (L)
DCW (L)
IR (L)
0.00mA~21.9mA
0.00mA~5.99mA
0001MΩ ~ 2000MΩ
Note
The L limit cannot be greater than the H limit.
When setting the current, be aware that a
maximum of 100VA can be set for ACW and 25W
for DCW.