Datasheet

D. E.
LOAD FUNCTION
The protection mechanism of OVP/OCP/OTP is implemented by
hardware circuit, which has the advantage of faster response time
than competitors who use software to achieve protection. When
it is detected that the voltage of the DUT exceeds the setting value
of the OVP, the output of the power supply can be stopped in a
short time to achieve the purpose of protecting the DUT.
Output delay function allows users to edit the timing waveform of
the power output on/off when the front panel voltage and current
settings are unchanged. In order to simplify the setting of waveform
editing, the GPP-Series has three built-in timing modes in the delay
output function, including Fixtime , Increase, Decline for users to
apply directly. The editing data of the output delay can be stored in
F.
OUTPUT DELAY FUNCTION
GPP-Series Delayed Waveform
OVP Trigger
GPP-Series Application
the internal 10 sets of memory, or to be saved by USB flash drive
(Save/Recall) and saved as *.DLY or *.CSV file. The stored *.CSV
can be exported into Excel for editing and analysis. The final
edited file can be exported to (Save/Recall) of the power supply
using a USB flash drive.
The output recorder function records the voltage & current
parameters of the output process. The recording interval of each
point can be set according to user's requirements, and the shortest
interval is 1 second and the longest is 300 seconds. The results can
be stored in *.REC or *.CSV format to the power supply or directly
G.
OUTPUT RECORDER FUNCTION
Schematic Diagram for Recorder Function Recorder Function Setting
Save as*.REC
saved in the USB flash drive. The stored *.CSV can be exported
into Excel to conduct the future analysis. (*.REC can be saved to
2018 records, *.CSV can be saved to 614400 records)
* Channel 3 of GPP-3323 does not support the output recorder function
The CH1/CH2 of the GPP-Series is designed with the load function. A
single power supply can meet the basic battery charging and discharging
test requirements. It can provide 32V/3A power output in channel 1 and
channel 2. The maximum 32V constant voltage load (CV), maximum 3.2A
constant current load (CC) and maximum 1k constant resistance load
(CR) function are built-in to allow users do conduct discharging test
without using an electronic load. In application, users can also set either
that one channel of the single GPP series as the power output, one
channel as the load function to consume the power of the DUT, or that
both channels as load functions to consume the power of different loads
.simultaneously