Instructions
GPM-8310 User Manual
240
The extended event register receives information about changes in the
condition register, which indicates the instrument’s internal condition.
The information is the result of edge detection performed by the
transition filter.
The following table lists the bit definitions for the condition register:
Bit
Name
Decimal
Definition
0
Updating
1
The measured data is being
updated. UPD changing from 1 to
0 indicates that updating has been
completed.
1
Integrate Busy
2
During integration.
2
Integrate Time Busy
4
The integration timer is operating.
3
Not Used
8
(Reserved for future use)
4
Frequency Over
16
The frequency is outside the
measurement range.
5
Store Busy
32
During storage.
6
Measured Data Over
64
The voltage or current exceeds its
range.
7
Voltage Peak Over
128
A peak over-range is detected in
the voltage.
8
Current Peak Over
256
A peak over-range is detected in
the current.
9
Not Used
512
(Reserved for future use)
10
Not Used
1024
(Reserved for future use)
11
Not Used
2048
(Reserved for future use)
12
Not Used
4096
(Reserved for future use)
13
Not Used
8192
(Reserved for future use)
14
Not Used
16384
(Reserved for future use)
15
Not Used
32768
(Reserved for future use)










