Instructions

GPM-8310 User Manual
212
Note
If :NUMeric:HOLD is set to ON before
:NUMeric[:NORMal]:VALue? or :NUMeric:LIST:VALue?
is executed, all the numeric data at that point in time
can be held internally.
As long as :NUMeric:HOLD is set to ON,numeric
data is held even when the numeric data on the screen
is updated.
If :NUMeric:HOLD is set to ON after having already
been set to ON before, the numeric data is cleared,
and the most recent numeric data is held internally.
When retrieving numeric data continuously, this
method can be used to circumvent the need to
repeatedly set
:NUMeric:HOLD to OFF.