Instructions
GPM-8310 User Manual
208
Description
Sets or returns the output components of the
harmonic measurement numeric list data.
Syntax
:NUMeric:LIST:SELect {EVEN|ODD|ALL}
Query Syntax
:NUMeric:LIST:SELect?
Parameter/
Return parameter
EVEN
Outputs the components of TOTal, DC, and
even-order harmonics.
ODD
Outputs the components of TOTal, DC, and
odd-order harmonics .
ALL
Outputs all components.
Example
:NUMERIC:LIST:SELECT ALL
:NUMERIC:LIST:SELECT?
->:NUMERIC:LIST:SELECT ALL
:NUMeric:LIST:ITEM<x>
Description
Sets or returns the output item (function and
element) of the specified harmonic measurement
numeric list data item.
Syntax
:NUMeric:LIST:ITEM<x>
{NONE|<Function>,<Element>}
Query Syntax
:NUMeric:LIST:ITEM<x>?
Parameter/
Return parameter
<x>
1 to 8 (item number)
NONE
No output item.
<Function>
{U|I|P|PHIU|PHII|UHDF|UHDF|PHDF
}
<Element>
1(If <Element> is omitted, the element
is set to 1)( For the GPM-8310, only set
to 1 or omitted)
Example
:NUMERIC:LIST:ITEM1 U,1
:NUMERIC:LIST:ITEM1?
->:NUMERIC:LIST:ITEM1 U,1
Set
Query