Instructions

GPM-8310 User Manual
196
Parameter/
Return parameter
ASCii
Select numeric data format to ascii.
FLOat
Select numeric data format to float.
Example
:NUMERIC:FORMAT ASCII
:NUMERIC:FORMAT?
->:NUMERIC:FORMAT ASCII
Note
ASCii
Physical values are output in the <NR3> format.
(Only the elapsed integration timeTIMEis
output in <NR1> format).
The data items are separated by commas.
FLOat
A header (for example, “#240” or “#3208”) is added
in front of each numeric data block.
A physical value in IEEE single-precision floating
point (4-byte) format follows the header.
#N (N-digit byte number)(data byte sequence).
The byte order of the data of each item is MSB First.
:NUMeric:NORMal
Description
Returns all normal numeric data output settings.
Query Syntax
:NUMeric:NORMal?
Return parameter
<String>
Note
The number of numeric data items output by :
NUMeric[:NORMal]:ITEM<x> is determined by :
NUMeric[:NORMal]:NUMber.
:NUMeric[:NORMal]:VALue
Description
Returns the numeric data.
Query Syntax
:NUMeric[:NORMal]:VALue? {<NRf>}
Query
Query