Instructions
COMMAND OVERVIEW
187
Return parameter
Bit0
The voltage is at the condition for
reducing the auto range or less.
Bit1
The voltage exceeds the condition for
raising the auto range.
Bit2
The voltage is over-range.
Bit3
The voltage is peak over-range.
Bit4
The current is at the condition for
reducing the auto range or less.
Bit5
The current exceeds the condition for
raising the auto range.
Bit6
The current is over-range.
Bit7
The current is peak over-range.
Example
:INPUT:CRANGE?
->:INPUT:CRANGE 8
(Indicate the voltage is peak over-range)
[:INPut]:ZERO
Description
Sets or returns the zero state.
Syntax
[:INPut]:ZERO {<Boolean>|OFF|ON}
Query Syntax
[:INPut]:ZERO?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the zero function off.
1
Turn the zero function on.
Example
:INPUT:ZERO OFF
:INPUT:ZERO?
->:INPUT:ZERO 0
Set
Query