Instructions
COMMAND OVERVIEW
163
Note
Example of a response with a header
:INPUT:VOLTAGE:RANGE 150.0E+00
Example of a response without a header
150.0E+00
:COMMunicate:LOCKout
Description
Sets or returns local lockout.
Syntax
:COMMunicate:LOCKout {<Boolean>|OFF | ON}
Query Syntax
:COMMunicate:LOCKout?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Disable the local key.
1
Enable the local key.
Example
:COMMUNICATE:LOCKOUT ON
:COMMUNICATE:LOCKOUT?
-> :COMMUNICATE:LOCKOUT 1
:COMMunicate:REMote
Description
Sets or returns the GPM-8310 series to remote or
local mode. ON is remote mode.
Syntax
:COMMunicate:REMote {<Boolean>|OFF | ON}
Query Syntax
:COMMunicate:REMote?
Set
Query
Set
Query