Instructions

GPM-8310 User Manual
160
Parameter/
Return parameter
<x>
<NRf>
1 to 4 (channel)
-9.999E+12~9.999E+12
Example
:AOUTPUT:RATE1 100,-100
:AOUTPUT:RATE1?
->:AOUTPUT:NORMAL:RATE1 100.0E+00,-100.0E+00
Note
When the D/A output is in manual range mode
Set the rated value for +5 V output and then that for
5 V output.
When the D/A output is in fixed range mode
There is no need to set these values. (The values do
not affect the output operation.)
:AOUTput:DIGital:MODE
Description
Sets the application mode of digital I/O (Remote
Control Only). For details, refer to page 130.
Syntax
:AOUTput:DIGital:MODE {IO|4094}
Query Syntax
:AOUTput:DIGital:MODE?
Parameter/
Return parameter
IO
4094
Sets the digital I/O to IO mode.
Sets the digital I/O to 4094 (serial to
parallel) mode.
Example
:AOUTPUT:DIGITAL:MODE IO
:AOUTPUT:DIGITAL:MODE?
->:AOUTPUT:DIGITAL:MODE IO
:AOUTput:DIGital:OUTPut
Description
When the 4094 mode (serial to parallel) is selected
for digital I/O, make use of this command to set
output status.
Syntax
:AOUTput:DIGital:OUTPut {<NR1>,<Boolean>}
Set
Query
Set