Instructions
COMMAND OVERVIEW
159
Parameter/
Return parameter
<x>
FIXed
MANual
1 to 4 (channel)
Fixed range mode.
Manual range mode.
Example
:AOUTPUT:NORMAL:MODE1 FIXED
:AOUTPUT:NORMAL:MODE1?
->:AOUTPUT:NORMAL:MODE1 FIXED
Note
• FIXed = Fixed range mode (default value)
Outputs +5 V when the rated value of each
measurement function is received.
• MANual = Manual range mode
The displayed values of the measurement function
when +5 V and –5 V are output as D/A output can
be set to any values of your choice. This enables the
D/A output to be expanded or reduced for each
channel (D/A zoom).
:AOUTput[:NORMal]:PRESet
Description
Sets the D/A output items to their default values.
Syntax
:AOUTput[:NORMal]:PRESet {NORMal|INTEGrate}
Parameter
NORMal
INTEGrate
Sets CH1:U, CH2:I, CH3:P, CH4:FU
Sets CH1:P, CH2:WH, CH3:AH, CH4:FU
Example
:AOUTPUT:NORAML:PRESET NORMAL
:AOUTput[:NORMal]:RATE<x>
Description
Sets or returns the maximum and minimum values
for when the D/A output is in manual range
mode. (Remote Control Only)
Syntax
:AOUTput[:NORMal]:RATE<x> {<NRf>,<NRf>}
Query Syntax
:AOUTput[:NORMal]:RATE<x>?
Set
Set
Query