Instructions

GPM-8310 User Manual
126
Manual (Manual
range mode)
You can set which measurement function values
result in a D/A output of -5V, and which result in a
D/A output of +5V. By doing so, you can enlarge or
reduce (zoom) the D/A output of each channel.
For example, if you are measuring a current that
fluctuates between 0.4A and 0.6A with a measurement
range of 1A, when the D/A output range mode is
Fixed, the D/A output voltage will fluctuate between
2.0V and 3.0V. When you want to observe the
fluctuations more closely, you can use the D/A zoom
feature. If you set the D/A output range mode to
Manual and set the minimum value to 0.4 and the
maximum value to 0.6, the instrument will produce
5V when the measured current value is 0.4A and +5V
when the measured current value is 0.6A.
Maximum and
Minimum Values
in Manual Range
Mode
The value maximum and minimum is between -
9.999E+12 and 9.999E+12.
The default value is 100.0 and -100.0.