Digital Power Meter GPM-8310 USER MANUAL Rev.
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company. The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice. Good Will Instrument Co., Ltd. No.
TABLE OF CONTENTS Table of Contents SAFETY INSTRUCTIONS .................................................. 5 GETTING STARTED ........................................................ 10 Characteristics .......................................... 11 Appearance ............................................... 16 Set Up ....................................................... 25 BASIC SETTING ............................................................. 28 Setting up measurement range .................
GPM-8310 User Manual Command List ........................................ 149 APPENDIX .................................................................... 228 Specifications ......................................... 229 Status system ......................................... 239 Dimensions ............................................ 243 Declaration of Conformity ....................... 244 Power measurement ............................... 245 Introduction to IEC-62301 .......................
SAFETY INSTRUCTIONS SAFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to ensure your safety and to keep the instrument in the best possible condition. Safety Symbols These safety symbols may appear in this manual or on the instrument. WARNING Warning: Identifies conditions or practices that could result in injury or loss of life.
GPM-8310 User Manual Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
SAFETY INSTRUCTIONS Safety Guidelines General Guideline Make sure that the voltage input level does not exceed AC600V. CAUTION Make sure the current input level does not exceed 20A. Do not place any heavy object on the instrument. Avoid severe impact or rough handling that can lead to damaging the instrument. Do not discharge static electricity to the instrument. Use only mating connectors, not bare wires, for the terminals. Do not block or obstruct the cooling fan vent opening.
GPM-8310 User Manual (Note) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The GPM-8310 falls under category II 600V. Measurement category IV is for measurement performed at the source of low-voltage installation. Measurement category III is for measurement performed in the building installation. Measurement category II is for measurement performed on the circuits directly connected to the low voltage installation.
SAFETY INSTRUCTIONS (Note) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The GPM-8310 falls under degree 2. Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”. Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. Pollution degree 2: Normally only non-conductive pollution occurs.
GPM-8310 User Manual GETTING STARTED This chapter describes the GPM-8310 in a nutshell, including accessories, package contents, its main features and front / rear panel introduction. Characteristics ................................................................ 11 Accessories .............................................................................................. 14 Package Contents .................................................................................... 15 Appearance ...............
GETTING STARTED Characteristics The GPM-8310 is a high-precision, programmable power meter for using in standby measuring the device with low power such as switching power supplies, transformers, power supplies, adapter and other devices. It equips with a color TFT-LCD screen and also multiple graph displays which are very convenient for reading the measurement results.
GPM-8310 User Manual Performance 12 6 selectable voltage ranges available from 15V to 600V with 0.1% of reading + 0.05% of range. 12 selectable current ranges available from 5mA to 20A with 0.1% of reading + 0.05% of range. It can even measure the voltage of abnormal wave of CF 3. The half-range CF is up to 6 or 6A. It can even measure the current of abnormal wave of CF 3. The half-range CF is up to 6 or 6A. Total harmonic distortion measurement.
GETTING STARTED Features Interface Application Full five-digit measurement. Voltage measurement range: 15V ~ 600V or automatic switching Current measurement range: 5mA ~ 20A or automatic switching Maximum accuracy of 0.1% of reading + 0.05% of range 2 main measurement readings and 8 minor measurement readings are displayed in the screen of standard display mode. 4 main measurement readings are displayed in the screen of simple display mode.
GPM-8310 User Manual Accessories Standard Accessories Part number 82PM-83100E01 Description User Manual CD 82GW1SAFE0M01 Safety Instruction Sheet Region dependent Power Cord GTL-209 Test leads: 1x red, 1x black GTL-212 Test leads: 1x yellow, 1x blue Optional Accessories Part number Option 14 Description GPM-001 Test Fixture GTL-234 RS232C cable GTL-246 USB cable GTL-248 GPIB cable GCP-300 Current Probe GRA-422 Rack Adapter Panel (19”, 2U) Name Description Opt.
GETTING STARTED Package Contents Check the contents before using the instrument.
GPM-8310 User Manual Appearance Front Panel Hardcopy Key LCD Display GPM - 8310 Function Keys Digital Power Meter Hardcopy V-Range I-Range MAX Hold Mode Setup Hold INTEGRATOR Start Stop Reset Key Lock Trigger ESC Local : Long Push POWER Enter Power key USB Host Port Soft keys Enter Key Arrow Keys Trigger Key ESC Key Power Key Turns On or Off the main power. For the power up sequence, see page 26. USB Host Port Connects with USB flash drive for data storage or screenshot.
GETTING STARTED Function Keys V-Range key, up/down arrow keys and Enter key can be used together to select a voltage range or auto range measurement mode. Also, press and hold the V-Range key to toggle between manual and auto range setting. See page 29. I-Range key, up/down arrow keys and Enter key can be used together to select a current range or auto range measurement mode. Also, press and hold the I-Range key to toggle between manual and auto range setting. See page 29.
GPM-8310 User Manual Press this key to toggle to key lock. In Remote control mode, press this button to switch to local mode. See page 90. Enter Key This button is used to enter the menu, confirm the settings and switch between the standard display mode and simple display mode (no function table and display icon). See page 89. Trigger key Activates the Trigger function. See page 89. ESC Key Press this button to cancel the current setting.
GETTING STARTED Main Display Overview Voltage Range Mode Sync Max. Hold Remote External Average Source Display Hold VT CT SF CF Update Rate Current Range Peak Voltage Peak Current Line Filter Harmonic Calculation Main Measurement Display Measure Storage Key Lock Minor Measurement Display Frequency Filter Remote Error Secondary Menus Item Status icon Description Voltage Range V-Range 15V Voltage measurement range. Example here range is 15V. V-Auto means that Voltage Auto Range is turned on.
GPM-8310 User Manual VT Ratio State VT External voltage magnification (On/Off) CT Ratio State CT External current magnification (On/Off) Power Ratio State SF External power magnification (On/Off) Remote RMT Remote control mode (On/Off) Line Filter L.F Voltage and current filters (On/Off) Frequency Filter F.F Frequency filters (On/Off) Maximum Hold MAX Hold Retain and display the maximum measurement reading. Average AVG-8 Average number of sampling (8/16/32/64) Sync Source SYNC.
GETTING STARTED Standard Display Display the measurement result of 2 major and 8 Mode minor measurement parameters Simple Display Mode Display the measurement result of 4 major measurement parameters Secondary menus Display secondary function menu Enlarge This function key is used to switch display of measurement result from 2 major plus 8 minor to 4 major ones. Integrator This function key is used to set up integrator measurement parameters and execute integrator measurement function.
GPM-8310 User Manual Rear Panel Current Input CURRENT I External Input 1/2 I Voltage Input V EXT1 RS-232 Connector LAN USB Device Port Port VOLTAGE RS232 LAN 10V MAX V EXT2 GPIB 600V MAX CAT II 600V Digital IO / DA4 WARNING GPIB Connector AC 100 240V 50 / 60Hz 30VA MAX 2V MAX Digital IO / DA4 Connector SER.NO. LABEL 20A MAX TO AVOID SHOCK, REMOVE INPUTS BEFORE OPENING. GND Terminal Power Cord Socket Power Cord Socket Connects the power cord.
GETTING STARTED LAN Port Accepts a LAN for remote control. For remote control details, see page 142. Digital IO / DA4 Connector Accepts a digital I/O cable for signal output; SCSI 26 pin, female connector. For digital I/O details, see page 122 GND Terminal Connects the GND (ground) terminal to the earth ground.
GPM-8310 User Manual Do not use damaged device. Before using the equipment, check its housing first to sure there is no any cracks. Do not operate this device in an environment containing explosive gases, steam or dust. The maximum measurable current and voltage are 600 V and 20A for voltage and current terminals of the rear panel of the GPM-8310. Do not input exceeded voltage and current, otherwise it will burn the device.
GETTING STARTED Set Up Tilting the Stand Pull out the handle sideways and rotate it clockwise for the several applications listed below. Horizontal Tilt Place the unit horizontally. Rotate the handle for tilt stand. Vertical Place the handle vertically for hand carry.
GPM-8310 User Manual Power Up Steps 1. Ensure the AC voltage is 100~ 240V. 2. Connect the power cord to the AC voltage input. Note Make sure the ground connector on the power cord is connected to a safety ground. This will influence the measurement accuracy. 3. Push to turn on the main power switch on the front panel. 4. The display turns on and shows the last function that was used before the power was reset.
GETTING STARTED Connect the wires to the GPM-8310 Two separate wires is used to connect the GPM-8310, so voltage and current measurement are isolated and don’t interfere with each other. Background Connection diagram CURRENT I I V EXT1 VOLTAGE 10V MAX V EXT2 20A MAX Description 2V MAX 600V MAX CAT II 600V V+ The positive voltage input (+), 600V for input on the rear panel. V- The negative voltage input (-), 600V for input on the rear panel.
GPM-8310 User Manual BASIC SETTING Setting up measurement range ........................................ 29 Auto Range .............................................................................................. 31 Setting up measurement status ....................................... 33 Setting up synchronization source ......................................................... 33 Setting up line filter .................................................................................
BASIC SETTING Setting up measurement range To get the accurate measurement results, you should set an appropriate measurement range before you perform measurement task. Set voltage range 1. Press V-Range button. The VRange field turns to bluish. 2. Use up and down arrow keys to select the desired range. 3. Press Enter button to confirm your selection. Available range Crest Factor AUTO, 15V, 30V, 60V, 150V, 300V, 600V is 3: Crest Factor AUTO, 7.5V, 15V, 30V, 75V, 150V, 300V is 6/6A: Set current range 1.
GPM-8310 User Manual 2. Use up and down arrow keys to select the desired range. 3. Press Enter button to confirm your selection. Available range Crest Factor AUTO, 5mA, 10mA, 20mA, 50mA, is 3: 100mA, 200mA, 0.5A, 1A, 2A, 5A, 10A, 20A Crest Factor AUTO, 2.5mA, 5mA, 10mA, 25mA, is 6/6A: 50mA, 100mA, 250mA, 0.5A, 1A, 2.5A, 5A, 10A Note Note Note 30 When it is under the manual setting ranging from 5mA to 200mA, the manual range adjusts to auto range automatically if input current exceeds 700mA.
BASIC SETTING Note The P.V status icon lights in red when the voltage measurement circuit detects that the measured value exceeds setting range by 3 folds (CF is set to 3) or 6 folds (CF is set to 6/6A). Auto Range The range is automatically switched according to the voltage and current of input signal. The range is shifted up when either of the following conditions is met. Range is shift up Vrms or Irms exceeds the measurement range by 130% at CF 3/6.
GPM-8310 User Manual Example To begin with, the measured Irms value is within the current range of I-Auto 20mA. The measured Irms (27.194mA) exceeds the IAuto 20mA by 130%, so the range is shifted up to 50mA automatically. The measured Irms (3.9994mA) is less than 30% of the I-Auto 20mA, so the range is shifted down to 10mA automatically.
BASIC SETTING Setting up measurement status Setting up synchronization source Steps 1. Press Setup button. 2. Press Enter button. 3. Press down arrow key to move cursor to the Sync Source field. 4. Use soft keys to select and confirm the desired option. Option Default value V Select the voltage of signals as synchronization source. The SYNC.V status icon, for example, on the display lights up in green when V is selected for sync source. I Select the current of signals as synchronization source.
GPM-8310 User Manual Setting up line filter Steps 1. Press Setup button. 2. Press Enter button. 3. Press down arrow key to move cursor to the Line Filter field. 4. Use soft keys to select and confirm the desired option. Option Default value 34 On Turn on the line filter function, which is inserted into voltage and current measurement input circuits and affects voltage, current as well as power measurements without high frequency components included within measured values. The L.
BASIC SETTING Setting up frequency filter Steps 1. Press Setup button. 2. Press Enter button. 3. Press down arrow key to move cursor to the Frequency Filter field. 4. Use soft keys to select and confirm the desired option. Option Default value On Turn on the frequency filter function, which is inserted into frequency measurement input circuit and affects frequency measurements with high frequency components included within measured values. The F.F status icon on the display lights up in green.
GPM-8310 User Manual Setting up crest factor Steps 1. Press Setup button. 2. Press Enter button. 3. Press down arrow key to move cursor to the Crest Factor field. 4. Use soft keys to select and confirm the desired option. Option Default value 36 3 Crest Factor is 3. 6 Crest Factor is 6. 6A Crest Factor is 6A where input range of measurement range will be extended and greater than 6.
BASIC SETTING Setting up auto-zero function Steps 1. Press Setup button. 2. Press Enter button. 3. Press down arrow key to move cursor to the Auto Zero field. 4. Use soft keys to select and confirm the desired option. Option Default value On Auto-zero function is activated once per hour or when range is switched. Off Auto-zero function is only activated once when the range is switched. The auto-zero function is turned off when the integrator function is executed.
GPM-8310 User Manual Setting up method of calculating harmonics Steps 1. Press Setup button. 2. Press Enter button. 3. Press down arrow key to move cursor to the Harmonics field. 4. Use soft keys to select and confirm the desired option. Option IEC Calculate the ratio of harmonic quantity of the 2nd through the upper limit 50th harmonic to the 1st harmonic. The HRM.I status icon, for example, on the display lights up in green when IEC is selected for harmonics.
BASIC SETTING 6. Use soft keys to increase or decrease the order number. Option 1-50 Default value 50 Set the upper limit of measured harmonic order within the range from 1 to 50.
GPM-8310 User Manual Setting up data update rate Steps 1. Press Setup button. 2. Press Enter button. 3. Press down arrow key to move cursor to the Data Update Rate field. 4. Use soft keys to select and confirm the desired option. Press the “More” soft keys to toggle among pages for further options. Option 0.1s/0.25s/ Measured value is updated in 0.5s/1s/2s/ accordance with the designated time 5s/10s/20s interval.
BASIC SETTING 6. Use soft keys to select and confirm the desired option. Option 1s/5s/ 10s/20s Default value 1s Note Time Out period acts like the time limit for detecting a period of the input waveform. Time Out function is only available when Auto is selected for Data Update Rate.
GPM-8310 User Manual Setting up measure storage Steps 1. Press Setup button. 2. Press Enter button. 3. Press down arrow key to move cursor to the Measure Storage field. Note Measure Storage function is Not available when Auto is selected for Data Update Rate. 4. Use soft keys to select and confirm the desired option. Option Default value Steps 42 On All measured date will be stored to the internal memory by designated time interval for repeating the storage operation.
BASIC SETTING 6. Use soft keys to increase or decrease the interval. Option The setting range for Interval is from 00:00:00 to 99:59:59. Default value 00:00:00 Note When it is set 00:00:00, the interval for measure storage will be synchronized with the designated Data Update Rate. Storage stops in the following circumstances: When data has been stored to all blocks, Normal measure data can be stored 10000 blocks and Normal with Harmonic data can be stored 1000 blocks.
GPM-8310 User Manual Setting up average function Steps 1. Press Setup button. 2. Press Average soft key. 3. Press Enter button. 4. Press down arrow key to move cursor to the State field. 5. Use soft keys to select and confirm the desired option. Option Default value Steps 44 On Turn Average function On for either Linear or Exponential averages of numeric data. It is particularly practical for large changes in load or power of low input signal frequency. Off Turn off Average function. Off 6.
BASIC SETTING 7. Use soft keys to select and confirm the desired option. Option Default value Steps Linear With the designated linear count, it is used to compute linear averages. Exponent With the specified attenuation count, numeric data will be averaged exponentially. Linear 8. Press down arrow key to move cursor to Count field. 9. Use soft keys to select and confirm the desired option.
GPM-8310 User Manual Setting up the voltage and current skipping configuration Steps 1. Press Setup button. 2. Press V / I Range soft key. 3. Press Enter button. 4. Press down arrow key to move cursor to the Mode field. 5. Use soft keys to select and confirm the desired option. Option Menu When user is configuring range setting, the measured data will Not be displayed. Quick The measured data will be displayed simultaneously while measurement range is being switched by user.
BASIC SETTING 7. Use soft keys to select and confirm the desired option. Option Default option Steps On It is able is skip certain measurement range(s) that are not used by turning on this feature. It can reduce measured data loss which happens while ranges are switched. Off Turn off the function. Off 8. Press down arrow key to move cursor to each field of both V-Range and I-Range. 9. Use soft keys to enable or disable the skipping function for each range.
GPM-8310 User Manual Default option Steps Off 10. Press down arrow key to move cursor to Peak Over field for VRange and I-Range, respectively. 11. Use soft keys to select and confirm the desired option. Press the More soft key to toggle among pages for Peak Over of V-Range and I-Range. Option When the occurrence of peak over-range happens in Auto range mode, user is able to define a measurement range to switch to. The available options for each mode are listed below.
BASIC SETTING Note The available options for Peak Over field are limited within the selected options from the VRange and I-Range sections above.
GPM-8310 User Manual Setting up the skipping configuration for external Steps 1. Press Setup button. 2. Press V / I Range soft key. 3. Press Enter button. 4. Press down arrow key to move cursor to Skipping Config field. 5. Use soft keys to select and confirm the desired option. Option Default option Steps On It is able is skip certain measurement range(s) that are not used by turning on this feature for external input. It can reduce measured data loss which occurs while ranges are switched.
BASIC SETTING 8. Press Enter button. 9. Press down arrow key to move cursor to each field of either External Sensor 1 or External Sensor 2. 10. Use soft keys to enable or disable the skipping function for each range. Option Default option Steps On The box of range will be checked when the range is enabled for skipping function. Off The range is disabled for skipping function. Off 11. Press down arrow key to move cursor to Peak Over field for External Sensor 1 or External Sensor 2, respectively.
GPM-8310 User Manual 12. Use soft keys to select and confirm desired option. Press More soft key to toggle among pages for Peak Over of Ext-1 and Ext-2, respectively. Option When the occurrence of peak over-range happens in Auto range mode for external input, user is able to define a measurement range to switch to. The available options for each mode are listed below. When it is under CF3 mode for External Sensor 1. Off/2.5V/5V/10V When it is under CF6/6A mode for External Sensor 1. Off/1.25V/2.
BASIC SETTING Setting up the VT ratio state Steps 1. Press Setup button. 2. Press Ratio soft key. 3. Press Enter button. 4. Press down arrow key to move cursor to the VT Ratio State field. 5. Use soft keys to select and confirm the desired option. Option Default option Steps On Turn on the VT (Voltage Transformer) ratio calculation function and the VT status icon on the display lights up in green. Off Turn off the VT ratio calculation function. Off 6.
GPM-8310 User Manual 7. Use soft keys to increase or decrease coefficient of VT ratio. Option The setting range for VT Ratio is from 0000.001 to 9999.999. Default value 0001.
BASIC SETTING Setting up the CT ratio state Steps 1. Press Setup button. 2. Press Ratio soft key. 3. Press Enter button. 4. Press down arrow key to move cursor to the CT Ratio State field. 5. Use soft keys to select and confirm the desired option. Option Default option Steps On Turn on the CT (Current Transformer) ratio calculation function and the CT status icon on the display lights up in green. Off Turn off the CT ratio calculation function. Off 6.
GPM-8310 User Manual 7. Use soft keys to increase or decrease coefficient of CT ratio. Option The setting range for CT Ratio is from 0000.001 to 9999.999. Default value 0001.
BASIC SETTING Setting up the power ratio state Steps 1. Press Setup button. 2. Press Ratio soft key. 3. Press Enter button. 4. Press down arrow key to move cursor to the Power Ratio State field. 5. Use soft keys to select and confirm the desired option. Option Default option Steps On Turn on the power ratio calculation function and the SF status icon on the display lights up in green. Off Turn off the power ratio calculation function. Off 6. Press down arrow key to move cursor to Ratio field.
GPM-8310 User Manual 7. Use soft keys to increase or decrease coefficient of power ratio. Option The setting range for power ratio is from 0000.001 to 9999.999. Default value 0001.
BASIC SETTING Setting up the external sensor input terminal Steps 1. Press Setup button. 2. Press External soft key. 3. Press Enter button. 4. Press down arrow key to move cursor to the External Sensor State field. 5. Use soft keys to select and confirm the desired option. Option Ext1 Turn on the Ext1 terminal function that receives voltage up to 10V including shunts and clamps from external output current sensor for measurement and the EXT1 status icon on the display lights up in green.
GPM-8310 User Manual Default option Steps Off 6. Press down arrow key to move cursor to either Ext1 Ratio (V/A) or Ext2 Ratio (mV/A) field. 7. Use soft keys to increase or decrease the conversion ratio of either Ext1 or Ext2. Option The setting range for both Ext1 and Ext2 is from 0000.001 to 9999.999. Default value Ext1 0001.000 Ext2 0010.000 Note 60 In order to enable range skipping configuration for external (page 50), it is required to enable external input function first.
BASIC SETTING Saving and loading the setup parameters Steps 1. Press Setup button. 2. Press Page 1/2 soft key. 3. Press Save Load soft key. 4. Press Enter button. 5. Press down arrow key to move cursor to the Type field. 6. Use soft keys to select and confirm the desired action. Option Default option Steps Save Select Save to store setup parameters into the internal memory. Load Select Load to recall setup parameters back from the internal memory. Save 7.
GPM-8310 User Manual 8. Use soft keys to select and confirm the desired memory set followed by clicking Ok soft key to confirm the Save or Load action. Option 1-4 Default option 1 62 There are 4 sets of internal memories for saving and loading setup parameters. The State field below indicates the status of selected memory set. Free represents the set is empty without saved parameters, whereas Saved indicates the set has been stored with setup parameters.
BASIC SETTING Setting up the D/A output configuration Steps 1. Press Setup button. 2. Press Page 1/2 soft key. 3. Press D/A soft key. 4. Press Enter button. Note Since the DA4 connector is an optional accessory, if it is not available on your unit, the D/A soft key will be disabled in grey color as the figure below shown. 5. Press down arrow key to move cursor to the Default Mode field.
GPM-8310 User Manual 6. Use soft keys to select and confirm the desired option. Option Normal The D/A output parameters for each channel will be changed to the default setting of Normal mode as follows. Normal Mode Integrator The D/A output parameters for each channel will be changed to the default setting of Integrator mode as follows. Integrator Mode Default option Steps 64 Default value CH1 V CH2 I CH3 P CH4 VHz Default value CH1 P CH2 WP CH3 q CH4 VHz Normal 7.
BASIC SETTING 8. Use soft keys to increase or decrease time for rated integrator. Option In the integrated values of D/A output, GPM-8310 presumes a rated value is received continuously over the designated time to be 100%, and assigns the value to 5V. The setting range for time of rated integrator is from 0000:00:00 to 9999:59:59. When the time is set 0000:00:00, D/A output value will be 0V. Default value 0001.00:00 Steps 9.
GPM-8310 User Manual I P VA VAR PF DEG VHz IHz VpK IpK WP WP+ WPq q+ qOff 66 Current Active power Apparent power Reactive power Power factor Phase angle Voltage frequency Current frequency Voltage peak Current peak Total watt hour Positive watt hour Negative watt hour Total ampere hour Positive ampere hour Negative ampere hour 0V D/A Output
BASIC SETTING Setting up the hardcopy and log configuration Steps 1. Press Setup button. 2. Press Page 1/2 soft key. 3. Press Hardcopy soft key. 4. Press Enter button. 5. Press down arrow key to move cursor to the Type field. 6. Use soft keys to select and confirm the desired option. Option Default option Steps Capture Select Capture to save screenshot file into the inserted USB disk. Log Select Log to save data log file into the inserted USB disk. Capture 7.
GPM-8310 User Manual 8. Use soft keys to select and confirm the desired action. Option Default option 68 On Turn on overwrite function so that the existed file within the USB disk will be overwritten when saving action is executed. Off By turning off overwrite function, a new saved file will be created and saved into the USB disk when executing saving action.
BASIC SETTING Setting up the MATH configuration Steps 1. Press Setup button. 2. Press Page 1/2 soft key. 3. Press MATH soft key. 4. Press Enter button. 5. Press down arrow key to move cursor to the Computation field. 6. Use soft keys to select and confirm the desired option.
GPM-8310 User Manual Steps 7. Press down arrow key to move cursor to Item A field. 8. Use soft keys to select and confirm the desired option.
BASIC SETTING Steps 9. Press down arrow key to move cursor to Item B field. 10. Use soft keys to select and confirm the desired option.
GPM-8310 User Manual Setting up System status System information screen Steps 1. Use left and right arrow keys on the front panel to select System function key. 2. Press Enter button to Enter SYSTEM INFORMATION screen where detailed information including Model, Serial Number, MCU/FPGA Version and MAC Address of the unit is displayed. 3. Press Enter button. 4. Press down arrow key to move cursor to Calibration Password field.
BASIC SETTING 5. Use soft keys along with left and right arrow keys to input the password followed by pressing Enter button twice to enter the Calibration page. Default option Note 99999 Refer to qualified technician and service manual for the calibration procedure. System configuration screen Steps 1. Use left and right arrow keys on the front panel to select System function key. 2. Press Enter button to Enter SYSTEM INFORMATION screen.
GPM-8310 User Manual 3. Press Config soft key to Enter SYSTEM CONFIG setting screen.
BASIC SETTING Setting up power on status Background Continue the following setting from SYSTEM CONFIG setting screen Steps 1. Press Enter button. 2. Press down arrow key to move cursor to Power On Status Setup field. 3. Use soft keys to select and confirm the desired option. Option Previous The status of unit on powering on is set to the status before the last shutdown. Default Default value The status of unit on powering on is set to the factory default status.
GPM-8310 User Manual Setting up brightness Background Continue the following setting from SYSTEM CONFIG setting screen Steps 1. Press Enter button. 2. Press down arrow key to move cursor to Brightness field. 3. Use soft keys to increase or decrease the brightness level Option 1 - 10 Default option 7 76 The display is the darkest when set to 1. On the contrary, it turns out the brightest when set to 10.
BASIC SETTING Setting up key sound Background Continue the following setting from SYSTEM CONFIG setting screen Steps 1. Press Enter button. 2. Press down arrow key to move cursor to Key Sound field. 3. Use soft keys to select and confirm the desired option. Option Default option On A short sound is heard from speaker of unit when pressing the keys on the front panel. Off No sound from speaker of unit when pressing the keys on the front panel.
GPM-8310 User Manual Setting up remote interface Background Continue the following setting from SYSTEM CONFIG setting screen Steps 1. Press Enter button. 2. Press down arrow key to move cursor to I/O Model field. 3. Use soft keys to select and confirm the desired option. Option 78 RS232 If interface is set to RS232, the Baud Rate and the Terminator fields can be selected. For details about configuring RS 232 interface, please see page 138.
BASIC SETTING USB For details about configuring USB interface, please see page 137. GPIB If interface is set to GPIB, the GPIB Address can be selected from “1” to “30”. Please see page 140 for details. LAN If interface is set to LAN, the IP model is can be selected from “Manual” and “DHCP”. For details about configuring LAN interface, please see page 142.
GPM-8310 User Manual Setting up SCPI identity Background Continue the following setting from SYSTEM CONFIG setting screen Steps 1. Press SCPI soft key to enter SCPI setting screen. 2. Press Enter button. 3. Press down arrow key to move cursor to Type field. 4. Use soft keys to select and confirm the desired option. Option Default The return message in remote control returns the default manufacturer, model number, serial number, among other info.
MEASUREMENT AND OTHER FUNCTIONS MEASUREMENT AND OTHER FUNCTIONS Measurement function .................................................... 82 Introduction to measurement parameters ............................................. 82 Setting measurement parameters .......................................................... 83 Changing the standard and simple display modes ................................ 86 Other functions...............................................................
GPM-8310 User Manual Measurement function The GPM-8310 provides a wide range of basic electricity and power measurement functions. It equips with different accurate measurement parameters for accurately measuring the voltage, current, power, DC/AC/AC + DC/V-MEAN, power factor, harmonics, frequency, etc. The input impedance of the device is 2MΩ, the maximum input voltage is 600Vrms. There are 2 sets of internal resistance (Shunt), 500mΩ and 5mΩ respectively.
MEASUREMENT AND OTHER FUNCTIONS Apparent Power VA Reactive power VAR Power Factor PF Phase Angle DEG Frequency IHz, VHz Voltage Peak V+pk, V-pk Current Peak I+pk, I-pk Active Power Peak P+pk, P-pk Total Harmonic Distortion THDI, THDV Crest factor CFV, CFI Mathematical Computation MATH Maximum Current Ratio MCR (Crest Factor(CFI) / Power Factor) Setting measurement parameters Steps 1. Use left and right arrow keys on the front panel to select Parameter function key.
GPM-8310 User Manual 2. Press Enter button. The 1st measurement parameter will be highlighted in green. 3. Press up, down, left and right arrow keys to select other desired measurement parameter.
MEASUREMENT AND OTHER FUNCTIONS 4. Press Enter button followed by using up and down arrow keys to switch display options for the selected measurement parameter. 5. User is able to apply the previous same process for each measurement parameter. There are up to 2 major and 8 minor measurement parameters to be switched.
GPM-8310 User Manual Changing the standard and simple display modes Steps 1. In the standard display mode, use left and right arrow keys on the front panel to select Enlarge function key. 2. Press Enter button to switch display to simple mode.
MEASUREMENT AND OTHER FUNCTIONS The simple mode covers 4 major measurement parameters deriving from the top 4 parameters of standard mode as shown below. 3. Press ESC button to return back to original display mode.
GPM-8310 User Manual Other functions Introduction to other functions Mode Function name 88 Button Max.
MEASUREMENT AND OTHER FUNCTIONS MAX Hold When the MAX Hold button is pressed, the MAX Hold status icon will light in red in the LCD display to indicate that this function is activated. To deactivate this function, press this button again. If the MAX Hold function is activated, the display value on the display is updated only when the current measured value is greater than the previous measured value. The maximum display value is retained on the display.
GPM-8310 User Manual ESC This button is used to exit current screen or return to the main measurement screen. Local/ Key Lock Dual function key. When Remote mode is activated, press this button to deactivate Remote mode and switch to Local mode. When Remote mode is not activated, this button is used as lock key of keypad. Mode Press the Mode button to select measurement mode. There are 4 measurement modes. 90 AC+DC: Displays all the components of measurement signal.
MEASUREMENT AND OTHER FUNCTIONS Integration measurement function Setting up Integrator measurement Steps 1. Use left and right arrow keys on the front panel to select Integrator function key. 2. Press Enter button to enter the integrator measurement screen.
GPM-8310 User Manual 3. Press right arrow key to move cursor to Set key. Select integrator measurement mode 92 4. Press Enter button to enter integrator measurement setting screen.
MEASUREMENT AND OTHER FUNCTIONS 5. Press Enter button to enter Mode field. Use up and down arrow keys to toggle between Manual, Standard and Continuous mode. Press Enter button again to confirm your selection. If you select Manual mode, the Set time become disable and displayed in gray. If you select Standard or Continuous mode, you need to set integrator measurement time before using integrator function. It can be set from 1 second to 9999 hours, 59 minutes and 59 seconds.
GPM-8310 User Manual Select integrator measurement function 6. Press down arrow key to move to Function field in the integrator measurement setting screen. 7. Press Enter button to enter Function field. Use up and down arrow keys to toggle between Ampere Hours and Watt Hours. Press Enter button again to confirm your selection.
MEASUREMENT AND OTHER FUNCTIONS If you select Ampere Hours, the measured value in the bottom half section will be displayed in “q”. If you select Watt Hours, the measured value in the bottom half section will be displayed in “WP”. Select integrator measurement parameter 8. Press down arrow key to move to the bottom half section where measured values are displayed.
GPM-8310 User Manual 9. Press Enter button to enter the 1st minor parameter followed by using up and down arrow keys to switch to preferred measurement parameter. Press Enter button again to confirm the selection. Press left or right arrow keys to move to the 2nd minor parameter followed by using the steps above to select a preferred parameter.
MEASUREMENT AND OTHER FUNCTIONS Introduction to integrator parameters Parameter name Description Mode Standard It allows user to define a period of Set Time for integrator measurement, which ranges from 1 second to 9999 hours, 59 minutes and 59 seconds. Manual User is not able to define a Set Time. The integrator measurement will be running constantly till Stop button is pressed by user.
GPM-8310 User Manual Set time It indicates the time of integrator measurement to be set. It can be set from 1 second to 9999 hours, 59 minutes and 59 seconds. State Running Integrator measurement is in progress. Stop Integrator measurement has been stopped manually. Timeout The time for running integrator measurement is up. Reset The integrator measurement status is cleared.
MEASUREMENT AND OTHER FUNCTIONS Measured value parameters For Watt Hours Positive total power: WP+ Negative total power: WPAverage power: P(avg) Voltage: Vdc (DC voltage), Vac (AC voltage), Vrms (AC+DC voltage),Vmn (Voltage mean) Current: Idc (DC current), Iac (AC current), Irms (AC+DC current) For Ampere Hours Total mAh: q Positive total mAh: q+ Negative total mAh: qAverage Current : q(avg) Voltage: Vdc (DC voltage), Vac (AC voltage), Vrms (AC+DC voltage), Vmn (Voltage mean) Current: Idc (DC current), Ia
GPM-8310 User Manual Using the integrator function Manual mode 1. In manual mode, you can directly press the Start button in the front panel to start integrator function. 2. To stop integration function, press the Stop button in the front panel. 3. Press the Reset button in the front panel to clear integrator.
MEASUREMENT AND OTHER FUNCTIONS Standard mode 1. Set integrator measurement time before using integrator function. 2. Other steps are same as running in manual mode. When integrator performing, the test time will increase until the setting integrator measurement time. Continuous mode 1. Set integrator measurement time before using integrator function. 2. Other steps are same as running in manual mode.
GPM-8310 User Manual Note 102 In the integration process, select the Measure key and press Enter button to return main measurement screen. Select Integrator key and press Enter button to switch back to integration measurement screen. In the integration process, you can Not change measurement range and enter system to set measurement parameters. In the integration process, if the voltage or current measurement value exceeds, the measured value will display in red.
MEASUREMENT AND OTHER FUNCTIONS Graph measurement function The GPM-8310 provides the professional graph measurement function via which user can have a well grip over fluctuations of measured values in waveform and harmonic in bar and list graphs in a friendly user interface. It is available, under the graph mode, to adjust both voltage and current ranges in real time and change the display modes along with relevant parameters with ease. Setting up waveform graph measurement Steps 1.
GPM-8310 User Manual 3. Press right arrow key to move cursor to Set key. Select waveform display mode 4. Press Enter button to enter waveform setting section. 5. Press Enter button to enter Display field. Use up and down arrow keys to toggle between options. Press Enter button again to confirm your selection.
MEASUREMENT AND OTHER FUNCTIONS Option Default option Select waveform display time division V, I, P Three items including the measured voltage, current and power are displayed in waveforms of different colors (V: yellow, I: red, Power: green) within the waveform chart. V, I Two items including the measured voltage and current are displayed in waveforms of different colors (V: yellow, I: red) within the waveform chart.
GPM-8310 User Manual 7. Press Enter button to enter Time Div field. Use up and down arrow keys to toggle between options. Press Enter button again to confirm your selection. Option 25us, 50us, 100us, 250us, 500us, 1ms, 2.5ms, 5ms, 10ms, 25ms, 50ms, 100ms, 250ms, 500ms, 1s Default option 5ms Note Select waveform display sync source 106 The diversified time units allow user to customize a preferred waveform graph display. In theory, shorter the measured period, smaller the time unit is fitting.
MEASUREMENT AND OTHER FUNCTIONS 9. Press Enter button to enter Sync field. Use up and down arrow keys to toggle between options. Press Enter button again to confirm your selection. Option Default option Select waveform display zoom magnification V Select the voltage of signals as synchronization source. I Select the current of signals as synchronization source. Off Select the entire interval of data updating period as synchronization source. V 10.
GPM-8310 User Manual 11. Press Enter button to enter Zoom (V) field. Use up and down arrow keys to toggle between options. Press Enter button again to confirm your selection. Option 1, 2, 3 The varied zoom magnifications allow user to customize a preferred waveform graph display. In theory, narrower the measured value, greater the zoom magnification is suitable. In contrast, wider the measured value, smaller the zoom magnification is fitting.
MEASUREMENT AND OTHER FUNCTIONS Frequency over limit When frequency of either voltage or current is beyond the limit, which varies per set Time Div, the warning message in the upper-right corner will be shown to alarm user as the figures below.
GPM-8310 User Manual Take few examples below that derive from the table above for further descroptions. When Update Time is set 20s, the range of Time Div is from 5ms to 1s and the available Frequency is up to 50Hz. When Update Time is set 0.1s, the range of Time Div is from 25us to 5ms and the available Frequency is up to 10kHz. Note The maximum frequency for Graph mode is up to10kHz.
MEASUREMENT AND OTHER FUNCTIONS Setting up waveform graph parameter Steps 1. Use left and right arrow keys on the front panel to select Graph function key. 2. Press Enter button to enter the Waveform graph display screen. 3. Press right arrow key to move cursor to Parameter key.
GPM-8310 User Manual 4. Press Enter button to enter parameters setting section. 5. Press Enter button to enter the 1st parameter. Use up and down arrow keys to toggle between options. Press Enter button again to confirm your selection. 6. Press down arrow key to move cursor to the 2nd parameter and repeat the above steps to set up. Also, repeat the same steps for the 3rd and 4th parameters.
MEASUREMENT AND OTHER FUNCTIONS Option Default option Voltage Vac (AC) Vdc (DC) Vrms (AC+DC) Vmn (V-MEAN) Current Iac (AC) Idc (DC) Irms (AC+DC, V-MEAN) Active Power P Apparent Power VA Reactive power VAR Power Factor PF Phase Angle DEG Frequency IHz, VHz Voltage Peak V+pk, V-pk Current Peak I+pk, I-pk Active Power Peak P+pk, P-pk Total Harmonic Distortion THDI, THDV Crest factor CFV, CFI Mathematical Computation MATH Maximum Current Ratio MCR Default options are based on the
GPM-8310 User Manual Setting up Harmonics bar graph measurement Steps 1. Use left and right arrow keys on the front panel to select Graph function key. 2. Press Enter button to enter the Waveform graph display screen.
MEASUREMENT AND OTHER FUNCTIONS 3. Press Enter button to enter Harmonics bar graph display screen where measured values of each harmonic order are shown in the histogram-like bar display. 4. Press right arrow key to move cursor to Set key. 5. Press Enter button to enter harmonics bar setting section.
GPM-8310 User Manual Select harmonics 6. Press Enter button to enter Display display mode Mode field. Use up and down arrow keys to toggle between options. Press Enter button again to confirm your selection. Option Default option V The THDV measured factor will be displayed in the right-side section in yellow and also shown in the left-side bar graph. I The THDI measured factor will be displayed in the right-side section in yellow and also shown in the left-side bar graph. V Select harmonics 7.
MEASUREMENT AND OTHER FUNCTIONS 8. Press Enter button to enter Order No. field. Use up and down arrow keys to toggle between options. Press Enter button again to confirm your selection. Option 1 - 50 Default option 1 Select a measured harmonic order with related values to be displayed in both the right-side section in green and the left-side bar graph. Note that the upper limit of order number is relevant to the Harmonics setting (page 38). Setting up Harmonics list graph measurement Steps 1.
GPM-8310 User Manual 2. Press Enter button to enter the Waveform graph display screen. 3. Press Enter button to enter Harmonics bar graph display screen. 4. Press right arrow key to move cursor to List key.
MEASUREMENT AND OTHER FUNCTIONS 5. Press Enter button to enter harmonics list display screen. Turn pages of harmonics list 6. Press right and left arrow keys to move cursor to Up and Down keys. Press Enter button for Up and Down keys individually to flip over pages of the harmonics list in which relevant values of each order of harmonics are well displayed. See the section below for descriptions of each item within the list.
GPM-8310 User Manual Items of the list 120 Order The harmonic order number V RMS voltage value of the harmonic order I RMS current value of the harmonic order P Active power value of the harmonic order V Hdf(%) Voltage harmonic distortion factor of the harmonic order I Hdf(%) Current harmonic distortion factor of the harmonic order P Hdf(%) Power harmonic distortion factor of the harmonic order V (°) The phase difference between the fundamental voltage and the voltage of the harmonic order
DIGITAL I/O / DA4 DIGITAL I/O / DA4 CURRENT I I V EXT1 VOLTAGE RS232 LAN 10V MAX V EXT2 GPIB 2V MAX 600V MAX CAT II 600V Digital IO / DA4 WARNING AC 100 240V 50 / 60Hz 30VA MAX SER.NO. LABEL 20A MAX TO AVOID SHOCK, REMOVE INPUTS BEFORE OPENING. Digital I/O / DA4 Overview ........................................... 122 External Remote Control ................................................ 124 DA4 Output Function .................................................... 125 User / 4094 Mode ...
GPM-8310 User Manual Digital I/O / DA4 Overview Background The digital I/O /DA4 port contains up to 3 modes: External Remote Control, DA4 Output Function and User-defined output function, which is divided into User Mode and 4094 Mode individually. Use the external I/O connector on the rear panel to control the instrument remotely and produce D/A output. By providing separate VCC power for the terminal, the outputs can also be used as a power source for TTL and CMOS circuits.
DIGITAL I/O / DA4 Note CAUTION 7 Out4/Serial Input 20 Digital GND 8 No connection 21 Digital GND 9 No connection 22 D/A ch4 out 10 D/A ch3 out 23 D/A ch2 out 11 D/A ch1 out 24 D/A GND 12 D/A GND 25 No connection 13 D/A GND 26 No connection The Digital GND and D/A GND signals are connected internally. 1. 2. 3. 4. 5. Do not apply voltage outside the range of 0 to 5 V to the remote control input pins. Also, do not short the output pins or apply external voltage to them.
GPM-8310 User Manual External Remote Control Overview Through external control, you can hold values, perform single measurements, and start, stop, and reset integration. Remote Control IO Circuit Input diagram Pulse width timing Start, Stop, Reset, hold, Trigger Output diagram Integ Busy Out Note 124 The Integ Busy output signal is set to low level during integration. Use this signal when you are observing integration.
DIGITAL I/O / DA4 DA4 Output Function Overview You can output voltage, current, active power, apparent power, reactive power, power factor, phase angle, frequency, voltage peak, current peak, and integrated values using a ±5V FS DC voltage. The output range mode and maximum/minimum value of manual range mode can only be used when using a remote control interface. Likewise this parameter can only be configured via remote control. Please see the commands on page 157 for full usage details.
GPM-8310 User Manual Manual (Manual You can set which measurement function values range mode) result in a D/A output of -5V, and which result in a D/A output of +5V. By doing so, you can enlarge or reduce (zoom) the D/A output of each channel. For example, if you are measuring a current that fluctuates between 0.4A and 0.6A with a measurement range of 1A, when the D/A output range mode is Fixed, the D/A output voltage will fluctuate between 2.0V and 3.0V.
DIGITAL I/O / DA4 Relationship between Output Items and D/A Output Voltage Frequency Integrator Other Examples of D/A Voltage: Output When the voltage range is set to 150 V and measurement value is 100 V, the output is 100 V/150 V × 5V = 3.3 V.
GPM-8310 User Manual Frequency: When the voltage frequency measurement value is 60Hz, the D/A output is 60 Hz/100 Hz × 5V = 3 V. Power: When the voltage range is set to 150 V and the current range is set to 2 A, the rated power range is 150 V × 2A = 300 W. When the measured power value is 150 W, the output is 150 W/300 W × 5V = 2.5 V. Integrated Power: When the voltage range is set to 150 V and the current range is set to 1 A, the rated power range is 150 V × 1A = 150 W.
DIGITAL I/O / DA4 Item Calculation V I P PF DEG VHz IHz (X / V_range) * 5V (X / I_range) * 5V (X / V_range * I_range) * 5V (X / V_range * I_range) * 5V (X / V_range * I_range) * 5V (X / 1.
GPM-8310 User Manual User / 4094 Mode Overview User (IO) and 4094 mode can only be used when using a remote control interface. Likewise this mode can only be enabled or disabled via remote control. Please see the digital I/O commands on page 160 for full usage details.
DIGITAL I/O / DA4 User Mode IO (Output) Mode Overview It is the mode utilizing output as general IO (Output) usage with up to 4 pins available for use simultaneously. Refer to the following introductions along with diagrams for more details. Please see the digital I/O commands on page 160 for full usage details.
GPM-8310 User Manual User Mode - Switch Mode (LED) Overview It is the mode driving LED as status display for user with up to 4 pins available for use simultaneously. Refer to the following introductions along with diagrams for more details. Please see the digital I/O commands on page 160 for full usage details.
DIGITAL I/O / DA4 User Mode - Switch Mode (Relay) Overview It is the mode driving Relay to control external circuit with up to 4 pins available for use simultaneously. Refer to the following introductions along with diagrams for more details. Please see the digital I/O commands on page 160 for full usage details.
GPM-8310 User Manual 4094 Mode Overview It is the mode for IO expansion via converting serial data into parallel data. Up to 8 pins are available simultaneously when single 4094 is in operation, whereas it rises to the maximum of 16 pins available simultaneously if putting two 4094 in series. Refer to the following introductions along with diagrams for more details. Please see the digital I/O commands on page 160 for full usage details.
DIGITAL I/O / DA4 Pin Diagram Use the built-in power supply Use the external power Method of series 135
GPM-8310 User Manual REMOTE CONTROL This chapter describes basic configuration of IEEE488.2 based remote control. For a command list, refer to the Command Overview chapter on page 145. Configure Remote Control Interface ............................... 137 Configure USB Interface .......................................................................... 137 Configure RS232 Interface ...................................................................... 138 Configure GPIB Interface .........................
REMOTE CONTROL Configure Remote Control Interface Configure USB Interface USB CDC Class Due to the USB port configured to CDC (Communications Device Class) by default, the GPM-8310 will appear as a virtual COM port to a connected PC. Before, hence, using remote control via CDC USB class, install the appropriate CDC USB driver included on the User Manual CD. Background Continue the following setting from SYSTEM CONFIG setting screen Steps 1. Press Enter button. 2.
GPM-8310 User Manual Configure RS232 Interface Background Steps Continue the following setting from SYSTEM CONFIG setting screen 1. Press Enter button. 2. Press down arrow key to move cursor to I/O Model field. 3. Use soft keys to select and confirm the RS232 option. 4. Press down arrow key to move cursor to Baud Rate field. 5. Use soft keys to select and confirm the Baud Rate option.
REMOTE CONTROL 6. Press down arrow key to move cursor to Terminator field. 7. Use soft keys to select and confirm the Terminator option. Option CR, LF, CR+LF Default value CR+LF RS232 Configuration Terminator indicates the end of line for return message.
GPM-8310 User Manual Configure GPIB Interface Background Continue the following setting from SYSTEM CONFIG setting screen Steps 1. Press Enter button. 2. Press down arrow key to move cursor to I/O Model field. 3. Use soft keys to select and confirm the GPIB option. 4. Press down arrow key to move cursor to GPIB Address field. 5. Use soft keys to increase or decrease to a target GPIB Address. Option 140 The range of GPIB Address is from 1 to 30.
REMOTE CONTROL Default option GPIB Pin Assignments 15 Pin Signal 1 Data I/O 1 2 Data I/O 2 3 Data I/O 3 4 Data I/O 4 5 EOI 6 DAV 7 NRFD Pin 13 14 15 16 17 18 19 8 NDAC 20 9 10 11 12 IFC SRQ ATN SHIELD Ground 21 22 23 24 Signal Data I/O 5 Data I/O 6 Data I/O 7 Data I/O 8 REN Ground (DAV) Ground (NRFD) Ground (NDAC) Ground (IFC) Ground (SRQ) Ground (ATN) Single GND 141
GPM-8310 User Manual Configure LAN Interface Background Continue the following setting from SYSTEM CONFIG setting screen Steps 1. Press Enter button. 2. Press down arrow key to move cursor to I/O Model field. 3. Use soft keys to select and confirm the LAN option. 4. Press down arrow key to move cursor to IP Model field. 5. Use soft keys to select and confirm the desired option. Option 142 Manual Set up IP Address, Subnet mask and Gateway manually.
REMOTE CONTROL DHCP Default option Steps DHCP server automatically assigns IP Address, Subnet mask and Gateway. DHCP 6. Press down arrow key to move cursor to Socket Port field. 7. Use soft keys to increase or decrease the parameter of Socket Port. Option The range of Socket Port is from 00000 to 65535. Default option 00023 Steps 8. When selecting Manual for IP Model, press down arrow key to move cursor to IP Address, Subnet Mask and Gateway fields, individually. 9.
GPM-8310 User Manual Configure EOL Character Description The system config menu can set the EOL(end-ofline) character for return message. (The USB, GPIB and LAN’s EOL character is fixed with CR+LF) The EOL characters that can be received from the PC include CR+LF, LF+CR, CR or LF. The most common EOL character is CR+LF. Return to Local Control Background Procedure When the unit is in remote control mode, the RMT icon above the main display can be seen.
COMMAND OVERVIEW COMMAND OVERVIEW The Command overview chapter lists all programming commands in functional order as well as alphabetical order. The command syntax section shows you the basic syntax rules you have to apply when using commands. Command Syntax Compatible Standard Command Structure Partial compatibility IEEE488.2 Partial compatibility SCPI, 1994 SCPI (Standard Commands for Programmable Instruments) commands follow a tree-like structure, organized into nodes.
GPM-8310 User Manual Command types Simple Example Query Command Forms A single command with/without a parameter :INPut:MODE DC A query is a simple or compound command followed by a question mark (?). A parameter (data) is returned. :INPut:CFACtor? Example Commands and queries have two different forms, long and short. The command syntax is written with the short form of the command in capitals and the remainder (long form) in lower case.
COMMAND OVERVIEW 300 :INPut:VOLTage:RANGe Command Format 2 1 1. 2. Command header Space Common Type Input Parameters 3. 3 Parameter 1 Description Example boolean logic 0, 1 integers 0, 1, 2, 3 decimal numbers 0.1, 3.14, 8.5 floating point with exponent 4.5e-1, 8.25e+1 any of NR1, 2, 3 1, 1.5, 4.5e-1 [MIN] (Optional parameter) For commands, this will set the setting to the lowest value.
GPM-8310 User Manual CR+LF Message Separator 148 EOL or ; Command Separator (semicolon) The most common EOL character is CR+LF
COMMAND OVERVIEW Command List *CLS....................................................................... 153 SCPI Commands *ESE ....................................................................... 153 *ESR ...................................................................... 154 *IDN ...................................................................... 154 *OPC ..................................................................... 154 *OPT ..............................................................
GPM-8310 User Manual :HARMonics:PLLSource ....................................... 171 :HARMonics:ORDer.............................................. 172 :HARMonics:THD ................................................. 172 HOLD Command :HOLD ................................................................... 173 :INPut .................................................................... 175 INPut Commands [:INPut]:WIRing ..................................................... 175 [:INPut]:MODE ................
COMMAND OVERVIEW :INTEGrate:STATe ................................................. 190 :MATH ................................................................... 191 Math commands MEASure commands NUMeric commands :MEASure ............................................................... 192 :MEASure:AVERaging ........................................... 192 :MEASure:AVERaging[:STATe] .............................. 192 :MEASure:AVERaging:TYPE .................................. 193 :MEASure:AVERaging:COUNt ..
GPM-8310 User Manual STATus commands STORe commands SYSTem commands 152 :STATus ................................................................. 217 :STATus:CONDition.............................................. 217 :STATus:EESE ........................................................ 217 :STATus:EESR........................................................ 218 :STATus:ERRor ...................................................... 218 :STATus:FILTer ................................................
COMMAND OVERVIEW SCPI Commands *CLS ...................................................................... 153 *ESE ...................................................................... 153 *ESR ...................................................................... 154 *IDN...................................................................... 154 *OPC ..................................................................... 154 *OPT .....................................................................
GPM-8310 User Manual *ESR Query Description Returns and clears the SESR (Standard Event Status Register). Query Syntax *ESR? Return parameter Example 0~255 *ESR? ->198 SESR=11000110 *IDN Query Description Returns the manufacturer, model number, serial number, and system version of the instrument. Query Syntax *IDN? Return parameter Example *IDN? ->GWInstek,GPM-8310, GXXXXXXXX,V1.
COMMAND OVERVIEW *OPT Query Description Returns the installed option. Query Syntax *OPT? Return parameter C1:GBIP C2:RS232 C3:USB Device C7:Ethernet EX1:External Sensor 1(2.
GPM-8310 User Manual Example *SER 7 Set the the SRER to 00000111 *SRE? ->3 SRER=00000011 *STB Query Description Returns the SBR (Status Byte Register) contents. Query Syntax *STB? Return parameter Example *STB ? ->34 SBR=00100010 *TRG Description 0~255 Set Executes single measurement (the same operation as when Trigger is pressed).
COMMAND OVERVIEW AOUTput Commands :AOUTput .............................................................. 157 :AOUTput[:NORMal]:CHANnel ..................... 157 :AOUTput[:NORMal]:IRTime ............................... 158 :AOUTput[:NORMal]:MODE .......................... 158 :AOUTput[:NORMal]:PRESet ............................... 159 :AOUTput[:NORMal]:RATE ............................ 159 :AOUTput:DIGital:MODE .................................... 160 :AOUTput:DIGital:OUTPut .....................
GPM-8310 User Manual Parameter/ 1 to 4 (channel) Return parameter U|I|P|S|Q|LAMBda|PHI|UPeak|IPeak|FU|F I|WH|WHP|WHM|AH|WHP|AHM|NONE 1(If is omitted, the element is set to 1)( For the GPM-8310, only set to 1 or omitted) Example :AOUTPUT:NORMAL:CHANNEL1 NONE Turns D/A channel1 output off (0V) :AOUTPUT:NORMAL:CHANNEL1? ->:AOUTPUT:NORMAL:CHANNEL1 I,1 Set :AOUTput[:NORMal]:IRTime Description Query Sets or returns the rated integration time that is used in the D/A output
COMMAND OVERVIEW Parameter/ Return parameter FIXed MANual Example 1 to 4 (channel) Fixed range mode. Manual range mode. :AOUTPUT:NORMAL:MODE1 FIXED :AOUTPUT:NORMAL:MODE1? ->:AOUTPUT:NORMAL:MODE1 FIXED Note • FIXed = Fixed range mode (default value) Outputs +5 V when the rated value of each measurement function is received. • MANual = Manual range mode The displayed values of the measurement function when +5 V and –5 V are output as D/A output can be set to any values of your choice.
GPM-8310 User Manual Parameter/ Return parameter Example 1 to 4 (channel) -9.999E+12~9.999E+12 :AOUTPUT:RATE1 100,-100 :AOUTPUT:RATE1? ->:AOUTPUT:NORMAL:RATE1 100.0E+00,-100.0E+00 Note • When the D/A output is in manual range mode Set the rated value for +5 V output and then that for –5 V output. • When the D/A output is in fixed range mode There is no need to set these values. (The values do not affect the output operation.
COMMAND OVERVIEW Parameter 0~255 (serial input data) 0,1 (strobe pulse) Example :AOUTPUT:DIGITAL:MODE 4094 :AOUTPUT:DIGITAL:OUTPUT 10,1 :AOUTput:DIGital:SETup Set Description When the IO mode is selected for digital I/O,make use of this command to set output status.
GPM-8310 User Manual COMMunciate Commands :COMMunicate ......................................................162 :COMMunicate:HEADer .......................................162 :COMMunicate:LOCKout ......................................163 :COMMunicate:REMote ........................................163 :COMMunicate:STATus ........................................164 :COMMunicate:VERBose ......................................164 :COMMunicate Query Description Returns all communication settings.
COMMAND OVERVIEW Note Example of a response with a header :INPUT:VOLTAGE:RANGE 150.0E+00 Example of a response without a header 150.0E+00 Set :COMMunicate:LOCKout Query Description Sets or returns local lockout. Syntax :COMMunicate:LOCKout {|OFF | ON} Query Syntax :COMMunicate:LOCKout? Parameter 0 OFF 1 ON Return parameter 0 1 Example Disable the local key. Enable the local key.
GPM-8310 User Manual Parameter 0 OFF 1 ON Return parameter 0 Turn the remote function off. 1 Turn the remote function on. Example :COMMUNICATE:REMOTE ON :COMMUNICATE:REMOTE? ->:COMMUNICATE:REMOTE 1 :COMMunicate:STATus Query Description Returns and clears the line-specific status.(Only for RS-232) Query Syntax :COMMunicate:STATus? Return parameter Bit 0 Parity error. (each status bit) Bit 1 Framing error. Bit 2 Noise error Break character detection.
COMMAND OVERVIEW Query Syntax :COMMunicate:VERBose? Parameter 0 OFF 1 ON Return parameter 0 Turn the verbose function off. 1 Turn the verbose function on. Example :COMMUNICATE:VERBOSE ON :COMMUNICATE:VERBOSE? ->:COMMUNICATE:VERBOSE 1 Note Example of a response fully spelled out :INPUT:VOLTAGE:RANGE 150.0E+00 Example of a response in abbreviated form :VOLT:RANG 150.
GPM-8310 User Manual DISPlay Commands :DISPlay ..................................................................166 :DISPlay:NORMal ..................................................166 :DISPlay[:NORMal]:ITEM ................................166 :DISPlay:INTegrate:ITEM ................................168 :DISPlay:PAGE .......................................................169 :DISPlay Query Description Returns all display settings.
COMMAND OVERVIEW Parameter/ 1 to 10 (display).
GPM-8310 User Manual ITHD Total harmonic distortion of current Ithd [THDI] MATH Mathematical Computation [MATH] MCR Maximum Current Ratio [MCR] Set :DISPlay:INTegrate:ITEM Query Description Sets or returns a Integrate measurement data display item. Refer to page 99 for details. Syntax :DISPlay:INTegrate:ITEM ,[,]} Query Syntax :DISPlay:INTegrate:ITEM? Parameter/ Return parameter Example 1 to 2(display).
COMMAND OVERVIEW Set :DISPlay:PAGE Query Description Sets or returns the display page item.
GPM-8310 User Manual HARMonics Command :HARMonics ..........................................................170 :HARMonics:DISPlay .............................................170 :HARMonics:DISPlay[:STATe] ...............................170 :HARMonics:DISPlay:ORDer ................................171 :HARMonics:PLLSource ........................................171 :HARMonics:ORDer ..............................................172 :HARMonics:THD .................................................
COMMAND OVERVIEW Parameter 0 OFF 1 ON Return parameter 0 Turn the harmionic display off. 1 Turn the harmionic display on. Example : HARMONICS:DISPLAY:STATE OFF : HARMONICS:DISPLAY:STATE? ->:HARMONICS:DISPLAY:STATE 0 Set :HARMonics:DISPlay:ORDer Query Description Sets or returns the harmonic order of the harmonic component that is shown in graph->hormoics->bar page for the harmonic measurement data display.
GPM-8310 User Manual Set :HARMonics:ORDer Description Query Sets or returns the maximum and minimum harmonic orders that are analyzed.
COMMAND OVERVIEW HOLD Command Set :HOLD Query Description Sets or returns the on/off state of the output hold feature for display, communication, and other types of data. Syntax :HOLD {|OFF|ON} Query Syntax :HOLD? Parameter 0 OFF 1 ON Return parameter 0 Turn the hold function off. 1 Turn the hold function on.
GPM-8310 User Manual INPut Commands :INPut .....................................................................175 [:INPut]:WIRing......................................................175 [:INPut]:MODE ......................................................176 [:INPut]:VOLTage ..................................................176 [:INPut]:VOLTage:RANGe .....................................176 [:INPut]:VOLTage:AUTO .......................................177 [:INPut]:VOLTage:CONFig ..........................
COMMAND OVERVIEW :INPut Query Description Returns all input settings. Query Syntax :INPut? Return parameter Set [:INPut]:CFACtor Query Description Sets or returns the crest factor. Syntax [:INPut]:CFACtor {3|6|A6} Query Syntax [:INPut]:CFACtor? Parameter/ 3 Return parameter 6 A6 crest factor 3. crest factor 6. Display range expand mode (6A) for crest factor 6.
GPM-8310 User Manual Set [:INPut]:MODE Query Description Sets or returns the voltage and current measurement mode. Syntax [:INPut]:MODE {DC|AC/RMS|ACDC| VMEan} Query Syntax [:INPut]:MODE? Parameter/ DC Return parameter AC/RMS Example Select the dc measurement mode. Select the ac measurement mode. ACDC Select the acdc measurement mode. VMEan Select the vmean measurement mode.
COMMAND OVERVIEW Example :INPUT:VOLTAGE:RANGE 600V :INPUT:VOLTAGE:RANGE? ->:INPUT:VOLTAGE:RANGE 600.0E+00 Set [:INPut]:VOLTage:AUTO Query Description Sets or returns the voltage auto range on/off state. Syntax [:INPut]:VOLTage:AUTO {|OFF|ON} Query Syntax [:INPut]:VOLTage:AUTO? Parameter 0 OFF 1 ON Return parameter 0 Turn the voltage auto range function off. 1 Turn the voltage auto range function on.
GPM-8310 User Manual Set [:INPut]:VOLTage:POJump Description Query Sets or returns the jump destination range that is used when a voltage peak over-range occurs. Syntax [:INPut]:VOLTage:POJump {OFF|} Query Syntax [:INPut]:VOLTage:POJump? Parameter/ OFF Return parameter Example No jump destination voltage range. See(:INPut:VOLTage:RANGe). :INPUT:VOLTAGE:POJUMP 600V :INPUT:VOLTAGE:POJUMP? ->:INPUT:VOLTAGE:POJUMP 600.
COMMAND OVERVIEW Parameter/ 1,2(EXT1,EXT2) Return parameter 5, 10, 20, 50, 100, 200, 500(mA) 1, 2, 5, 10, 20(A) when the crest factor is set to 3. 2.5, 5, 10, 25, 50, 100, 250(mA) 0.5, 1, 2.5, 5, 10(A) when the crest factor is set to 6 or 6A. EXTernal1 2.5, 5, 10(V) when the crest factor is set to 3. 1.25, 2.5, 5(V) when the crest factor is set to 6 or 6A. EXTernal2 50, 100, 200, 500,(mV), 1, 2(V) when the crest factor is set to 3. 25, 50, 100, 250,(mV), 0.
GPM-8310 User Manual Example :INPUT:CURRENT:AUTO ON :INPUT:CURRENT:AUTO? ->:INPUT:CURRENT:AUTO 1 Set [:INPut]:CURRent:CONFig Query Description Sets or returns the valid current range. Syntax [:INPut]:CURRent:CONFig {ALL|[,Current]…} Query Syntax [:INPut]:CURRent:CONFig? Parameter/ ALL Return parameter Example All ranges are valid. See(:INPut:CURRent:RANGe). :INPUT:CURRENT:CONFIG 20,10,1 :INPUT:CURRENT:CONFIG? ->:INPUT:CURRENT:CONFIG 20.0E+00,10.0E+00, 1.
COMMAND OVERVIEW Set [:INPut]:CURRent:EXTSensor:CONFig Description Query Sets or returns the valid external current sensor range. Syntax [:INPut]:CURRent:EXTSensor:CONFig {ALL|[,Voltage]…} Query Syntax [:INPut]:CURRent:EXTSensor:CONFig? Parameter/ Return parameter Example 1,2(EXT1,EXT2), If is omitted, by default sets or returns EXT2 config. ALL All ranges are valid. See(:INPut:CURRent:RANGe). :INPUT:CURRENT:EXTSENSOR:CONFIG1 2,0.5,0.
GPM-8310 User Manual Set [:INPut]:CURRent:SRATio:ELEMent1 Query Description Sets or returns the external current sensor conversion ratio of the specified element. Syntax [:INPut]:CURRent:SRATio:ELEMent1 {} Query Syntax [:INPut]:CURRent:SRATio:ELEMent1? Parameter/ Return parameter Example 1,2(EXT1,EXT2), If is omitted, by default sets or returns EXT2 config. 0.
COMMAND OVERVIEW Set [:INPut]:RCONfig Query Description Sets or returns the on/off state of the range configuration (valid range selection) feature. Syntax [:INPut]:RCONfig {|OFF|ON} Query Syntax [:INPut]:RCONfig? Parameter 0 OFF 1 ON Return parameter 0 Turn the range configuration feature off. 1 Turn the range configuration feature on. Example :INPUT:RCONFIG ON :INPUT:RCONFIG? ->:INPUT:RCONFIG 1 [:INPut]:SCALing Query Description Returns all scaling settings.
GPM-8310 User Manual Example :INPUT:SCALING:STATE ON :INPUT:SCALING:STATE? ->:INPUT:SCALING:STATE 1 [:INPut]:SCALing:{VT|CT|SFACtor}:ELEMen t Set Query Description Sets or returns the VT ratio, CT ratio, or power coefficient of the specified element.
COMMAND OVERVIEW Example :INPUT:SYNCHRONIZE VOLTAGE :INPUT:SYNCHRONIZE? ->:INPUT:SYNCHRONIZE VOLTAGE [:INPut]:FILTer Query Description Returns all input filter settings. Query Syntax [:INPut]:FILTer? Return parameter Set [:INPut]:FILTer:LINE Query Description Sets or returns the line filter. Syntax [:INPut]:FILTer:LINE {|OFF|ON} Query Syntax [:INPut]:FILTer:LINE? Parameter 0 OFF 1 ON Return parameter 0 Turn the line filter function off.
GPM-8310 User Manual Parameter 0 OFF 1 ON Return parameter 0 Turn the frequency filter function off. 1 Turn the frequency filter function on. Example :INPUT:FILTER:FREQUECNY OFF :INPUT:FILTER:FREQUECNY? ->:INPUT:FILTER:FREQUECNY 0 [:INPut]:POVer Query Description Returns the peak over-range information. Query Syntax [:INPut]:POVer? Return parameter Bit0 Voltage peak over-range is occurring. Bit1 Current peak over-range is occurring.
COMMAND OVERVIEW Return parameter Bit0 The voltage is at the condition for reducing the auto range or less. Bit1 The voltage exceeds the condition for raising the auto range. Bit2 The voltage is over-range. Bit3 The voltage is peak over-range. Bit4 The current is at the condition for reducing the auto range or less. Bit5 The current exceeds the condition for raising the auto range. Example Bit6 The current is over-range. Bit7 The current is peak over-range.
GPM-8310 User Manual INTEGrate Commands :INTEGrate .............................................................188 :INTEGrate:MODE ................................................188 :INTEGrate:FUNCtion ...........................................189 :INTEGrate:TIMer ..................................................189 :INTEGrate:STARt ..................................................189 :INTEGrate:STOP ..................................................190 :INTEGrate:RESet ...............................
COMMAND OVERVIEW Set :INTEGrate:FUNCtion Query Description Sets or returns the integration function. Syntax :INTEGrate:FUNCtion {WATT|AMPEre} Query Syntax :INTEGrate:FUNCtion? Parameter/ WATT Return parameter AMPEre Example Select the integration function watt. Select the integration function ampere. :INTEGRATE:FUNCTION WATT :INTEGRATE:FUNCTION? ->:INTEGRATE:FUNCTION WATT Set :INTEGrate:TIMer Query Description Sets or returns the integration timer value.
GPM-8310 User Manual :INTEGrate:STOP Set Description Stops integration. Syntax :INTEGrate:STOP Example :INTEGRATE:STOP :INTEGrate:RESet Set Description Resets the integrated value. Syntax :INTEGrate:RESet Example :INTEGRATE:RESET :INTEGrate:STATe Query Description Returns the integration status. Query Syntax :INTEGrate:STATe? Return parameter ERRor Example RESet Integration resets. STARt Integration is in progress. STOP Integration stops.
COMMAND OVERVIEW Math Commands Set :MATH Query Description Sets or returns the MATH equation.
GPM-8310 User Manual MEASure Commands :MEASure ...............................................................192 :MEASure:AVERaging ............................................192 :MEASure:AVERaging[:STATe] ..............................192 :MEASure:AVERaging:TYPE ..................................193 :MEASure:AVERaging:COUNt ..............................193 :MEASure:MHOLd ................................................
COMMAND OVERVIEW Example :MEASURE:AVERAGING:STATE ON :MEASURE:AVERAGING:STATE? ->:MEASURE:AVERAGING:STATE 1 Set :MEASure:AVERaging:TYPE Query Description Sets or returns the averaging type. Syntax :MEASure:AVERaging:TYPE {LINear|EXPonent} Query Syntax :MEASure:AVERaging:TYPE? Parameter/ LINear Return parameter EXPonent Example Select averaging type to linear. Select averaging type to exponent.
GPM-8310 User Manual Parameter 0 OFF 1 ON Return parameter 0 Turn the MAX hold function off. 1 Turn the MAX hold function on.
COMMAND OVERVIEW NUMeric Commands :NUMeric .............................................................. 195 :NUMeric:FORMat ............................................... 195 :NUMeric:NORMal ............................................... 196 :NUMeric[:NORMal]:VALue ................................. 196 :NUMeric[:NORMal]:NUMber ............................. 198 :NUMeric[:NORMal]:ITEM ............................. 198 :NUMeric[:NORMal]:PRESet ................................
GPM-8310 User Manual Parameter/ ASCii Return parameter FLOat Example Select numeric data format to ascii. Select numeric data format to float. :NUMERIC:FORMAT ASCII :NUMERIC:FORMAT? ->:NUMERIC:FORMAT ASCII Note • ASCii Physical values are output in the format. (Only the elapsed integration time—TIME—is output in format). The data items are separated by commas. • FLOat A header (for example, “#240” or “#3208”) is added in front of each numeric data block.
COMMAND OVERVIEW Parameter 1 to 50 (item number) Example • If is specified, only the numeric data for the specified item is output. :NUMERIC:NORMAL:VALUE? 1 -> 103.79E+00 • If is omitted, the numeric data items from 1 to the number specified by the : NUMeric[:NORMal]:NUMber command are output in order. :NUMERIC:NORMAL:VALUE? -> 103.79E+00,1.0143E+00,105.27E+00,..(omitted)..,50.
GPM-8310 User Manual Error Data Data does not exist (the display shows “-----”) ASCII: NAN (Not A Number) FLOAT: 0x7E951BEE (9.91E+37) Data over (the display shows “-----“) ASCII: INF (INFinity) FLOAT: 0x7E94F56A (9.9E+37) Set :NUMeric[:NORMal]:NUMber Description Query Sets or returns the number of numeric data items that are transmitted by the :NUMeric[:NORMal]: VALue? command.
COMMAND OVERVIEW Parameter/ Return parameter NONE 1 to 50 (item number) No output item.
GPM-8310 User Manual PPPeak Maximum power: P+pk [P+pk] PMPeak Minimum power: P-pk [P-pk] S Apparent power S [VA] Q Reactive power Q [VAR] LAMBda Power factor λ [PF] CFU Voltage factor λ [CFV] CFV Current factor λ [CFI] PHI Phase difference Φ [DEG] FU Voltage frequency fu [VHz] FI Current frequency fI [AHz] UTHD Total harmonic distortion of voltage Uthd [THDV] ITHD Total harmonic distortion of current Ithd [THDI] WH Watt hour WP [WP] WHP Positive watt hour WP+ [WP+]
COMMAND OVERVIEW URMN Rectified mean voltage Urmn UAC AC voltage component Uac IRMS True rms current Irms IMN Rectified mean current calibrated to the rms value Imn IDC Simple current average Idc IRMN Rectified mean current Irmn IAC AC current component Iac [Iac] UK Rms voltage of harmonic order k U(k) [V] IK Rms current of harmonic order k I(k) [A] PK Active power of harmonic order k P(k) [P] LAMBDAK Power factor of harmonic order k λ(k) PHIK Phase difference between the voltage
GPM-8310 User Manual PHDFk Harmonic distortion factor of power Phdf(k) :NUMeric[:NORMal]:PRESet Set Description Presets the numeric data output item pattern.
COMMAND OVERVIEW Patterns 4 6 LAMBda 7 PHI 8 FU 9 FI 10 UPPeak 11 UMPeak 12 IPPeak 13 IMPeak 14 PPPeak 15 PMPeak ITEM 1 U 2 I 3 P 4 S 5 Q 6 LAMBda 7 PHI 8 FU 9 FI 10 UPPeak 11 UMPeak 12 IPPeak 13 IMPeak 14 TIME 15 WH 16 WHP 17 WHM 203
GPM-8310 User Manual 18 AH 19 AHP 20 AHM :NUMeric[:NORMal]:CLEar Set Description Clears numeric data output items (sets the items to NONE). Syntax :NUMeric[:NORMal]:CLEar {ALL|[,]} Parameter ALL Clear all items.
COMMAND OVERVIEW Note When output items are deleted, subsequent items shift forward to fill the empty positions. Empty positions at the end are set to NONE. If the second is omitted, only the output item specified by the first number is deleted. :NUMeric[:NORMal]:HEADer Query Description Returns the numeric data header. Syntax :NUMeric[:NORMal]:HEADer? {} Parameter Example • If is specified, only the data name for the specified item number is output.
GPM-8310 User Manual :NUMeric:LIST:VALue Query Description Returns the harmonic measurement numeric list data. Query Syntax :NUMeric:LIST:VALue? {} Parameter Example • if is specified :NUMERIC:LIST:VALUE? 1 1 to 8 (item number) -> 103.58E+00,NAN,103.53E+00,0.09E+00,2.07E+00, 0.04E+00,.. (omitted)..,0.01E+00,0.01E+00 (up to 52 data values) • if is omitted (when :NUMeric:LIST:NUMber is set to 5) :NUMERIC:LIST:VALUE? -> 103.58E+00,NAN,103.53E+00,0.09E+00,2.07E+00, 0.04E+00,..
COMMAND OVERVIEW Description Sets or returns the number of numeric list data items that are transmitted by :NUMeric:LIST:VALue? command. Syntax :NUMeric:LIST:NUMber {|ALL} Query Syntax :NUMeric:LIST:NUMber? Parameter/ Return parameter Example 1 to 8(ALL) :NUMERIC:LIST:NUMBER 3 :NUMERIC:LIST:NUMBER? ->:NUMERIC:LIST:NUMBER 3 Note If the parameter is omitted from the :NUMeric:LIST:VALue? command, the numeric list data items from 1 to the specified value are output in order.
GPM-8310 User Manual Description Sets or returns the output components of the harmonic measurement numeric list data. Syntax :NUMeric:LIST:SELect {EVEN|ODD|ALL} Query Syntax :NUMeric:LIST:SELect? Parameter/ EVEN Return parameter Outputs the components of TOTal, DC, and even-order harmonics. ODD Outputs the components of TOTal, DC, and odd-order harmonics . ALL Outputs all components.
COMMAND OVERVIEW :NUMeric:LIST:PRESet Description Set Presets the harmonic measurement numeric list data output item pattern.
GPM-8310 User Manual 3 P 4 PHIU 5 PHII 6 UHDF 7 IHDF 8 PHDF :NUMeric:LIST:CLEar Set Description Clears numeric data output items (sets the items to NONE). Syntax :NUMeric:LIST:CLEar {ALL|[,]} Parameter ALL Clear all items.
COMMAND OVERVIEW Example :NUMERIC:LIST:DELETE 1 (Deletes ITEM1 and shifts ITEM2 and subsequent items forward). :NUMERIC:LIST:DELETE 1,3 (Deletes ITEM1 to ITEM3 and shifts ITEM4 and subsequent items forward). Note When output items are deleted, subsequent items shift forward to fill the empty positions. Empty positions at the end are set to NONE. If the second is omitted, only the output item specified by the first number is deleted.
GPM-8310 User Manual Note • If :NUMeric:HOLD is set to ON before :NUMeric[:NORMal]:VALue? or :NUMeric:LIST:VALue? is executed, all the numeric data at that point in time can be held internally. • As long as :NUMeric:HOLD is set to ON,numeric data is held even when the numeric data on the screen is updated. • If :NUMeric:HOLD is set to ON after having already been set to ON before, the numeric data is cleared, and the most recent numeric data is held internally.
COMMAND OVERVIEW RATE Commands :RATE ..................................................................... 213 :RATE:AUTO ......................................................... 213 :RATE:AUTO:TIMeout .......................................... 213 :RATE:AUTO:SYNChronize .................................. 214 Set :RATE Query Description Sets or returns the data update interval.
GPM-8310 User Manual Query Syntax :RATE:AUTO:TIMeout? Parameter/
COMMAND OVERVIEW RECall Commands :RECall:NUMber ................................................... 215 :RECall[:NORMal]:VALue ...................................... 215 :RECall:LIST:VALue ............................................... 216 :RECall:PANel........................................................ 216 :RECall:NUMber Description Query Returns the number of blocks of measured data that is stored.
GPM-8310 User Manual :RECall:LIST:VALue Description Query Returns the numeric list data of harmonic measurement at the specified block number. Query Syntax :RECall:LIST:VALue? {} Parameter Example • if is specified,the numeric list data at the specified block number will be returned.
COMMAND OVERVIEW STATus Commands :STATus ................................................................. 217 :STATus:CONDition ............................................. 217 :STATus:EESE ....................................................... 217 :STATus:EESR ....................................................... 218 :STATus:ERRor ...................................................... 218 :STATus:FILTer ............................................... 219 :STATus:QENable ...........................
GPM-8310 User Manual Query Syntax :STATus:EESE? Parameter/ Return parameter Example 0 to 65535 :STATUS:EESE 16 :STATUS:EESE? -> :STATUS:EESE 16 Note For information about the condition register, see Appendix,"Status system" at page 239. :STATus:EESR Query Description Returns the contents of the extended event register and clears the register.
COMMAND OVERVIEW Note • If no errors have occurred, 0,”No error” is returned. • User can use the :STATus:QMESsage command to specify whether the message is included. • Error message description: Error_103: Invalid separator Error_104: Data type error. Error_108: Parameter not allowed. Error_109: Missing parameter. Error_113: Undefined header. Error_131: Invalid suffix. Error_141: Invalid character data. Error_221: Setting conflict. Error_222: Data out of range. Error_813: Invalid operation.
GPM-8310 User Manual Note • Set how each bit in the condition register must change to trigger the setting of an event. • For information about the condition register, see Appendix,"Status system" at page 239. Set :STATus:QENable Query Description Sets or returns whether messages other than errors will be stored to the error queue (ON) or not (OFF). Syntax :STATus:QENable {|OFF|ON} Query Syntax :STATus:QENable? Parameter 0 OFF 1 ON Return parameter 0 Function is off.
COMMAND OVERVIEW Example :STATUS:QMESSAGE ON :STATUS:QMESSAGE? -> :STATUS:QMESSAGE 1 221
GPM-8310 User Manual STORe Commands :STORe ...................................................................222 :STORe[:STATe] .....................................................222 :STORe:INTerval ....................................................222 :STORe:PANel ........................................................223 :STORe Query Description Returns all storage settings.
COMMAND OVERVIEW Parameter/ 1st Return parameter 2nd 3rd Example 0 to 99 (hours) 0 to 59 (minutes) 0 to 59 (seconds) :STORE:INTERVAL 0,0,1 :STORE:INTERVAL? ->:STORE:INTERVAL 0,0,1 Note • When time interval is set 00:00:00, the storage interval is identical with the designated data update interval. :STORe:PANel Set Description Saves setup parameters to a file.
GPM-8310 User Manual SYSTem Commands :SYSTem .................................................................224 :SYSTem:BRIGhtness ............................................224 :SYSTem:COMMunicate:COMMand ....................224 :SYSTem:COMMunicate:ETHernet:MACaddress 225 :SYSTem:FIRMware:DATE ....................................225 :SYSTem:KLOCk ....................................................226 :SYSTem:MODel....................................................226 :SYSTem:RESolution ..............
COMMAND OVERVIEW Description Sets or returns the command type. Syntax :SYSTem:COMMunicate:COMMand {DEFAULT|USER} Query Syntax :SYSTem:COMMunicate:COMMand? Parameter/ DEFAULT Return parameter USER Example GPM8310. User-define. :SYSTEM:COMMUNICATE:COMMAND DEFAULT :SYSTEM:COMMUNICATE:COMMAND? ->:SYSTEM:COMMUNICATE:COMMAND DEFAULT Note • The SCPI mode is used to determine whether the *IDN? query returns the “Default” or “User” identification string.
GPM-8310 User Manual Syntax :SYSTem:KEY:BEEPer {|OFF|ON} Query Syntax :SYSTem:KEY:BEEPer? Parameter 0 OFF 1 ON Return parameter 0 Turn the keyclick beeper function off. 1 Turn the keyclick beeper function on. Example :SYSTEM:KEY:BEEPER OFF :SYSTEM:KEY:BEEPER? ->:SYSTEM:KEY:BEEPER 0 Set :SYSTem:KLOCk Query Description Sets or returns the on/off state of the key protection.
COMMAND OVERVIEW :SYSTem:RESolution Query Description Returns the numeric data display resolution. Query Syntax :SYSTem:RESolution? Example :SYSTEM:RESOLUTION? ->:SYSTEM:RESOLUTION 5 :SYSTem:SERial Description Returns the serial number. Syntax :SYSTem:SERial? Example :SYSTEM:SERIAL? Query ->:SYSTER:SERIAL 123456789A :SYSTem:VERsion[:FIRMware] Description Returns the firmware version. Query Syntax :SYSTem:VERsion[:FIRMware]? Example :SYSTEM:VERSION:FIRMWARE? Query ->"V1.
APPENDIX APPENDIX Specifications ............................................................... 229 General Specifications ............................................................................. 229 Input ........................................................................................................ 230 Voltage and Current Accuracy ................................................................. 231 Active Power Accuracy .....................................................................
APPENDIX Specifications Below are the basic conditions required to operate the GPM-8310 within specification: Calibration: Yearly Operating Environment: 18~28 ˚C (64.4~82.4˚F) Humidity: <80%RH, Accuracy: ± (% of reading + % of range) The specifications apply when it warmed up for at least 30 minutes and operates in the slow rate. The power supply cable must be grounded to ensure accuracy. Input voltage and current must be standard sine wave. The power factor must be 1.
GPM-8310 User Manual Input Item Input type Measure range Input impedance Continuous maximum allowable input Input bandwidth Continuous maximum Common-mode voltage Line filter Frequency filter A/D converter 230 Specifications Voltage Floating input through resistive voltage divider Current Floating input through shunt Voltage 15 V, 30 V, 60 V,150 V, 300 V, 600 V Current Direct 5 mA, 10 mA, 20 mA, 50 mA, 100 mA, input 200 mA, 0.5 A, 1 A, 2 A, 5 A, 10 A, 20 A Sensor EX1: 2.
APPENDIX Voltage and Current Accuracy Item Requirements Accuracy Temperature coefficient When the line filter is turned ON Specifications Temperature Humidity Input waveform common-mode voltage Number of displayed digits Frequency filter 23 ± 5℃ 30~75% RH Sine wave crest factor = 3 0V 5 digits Turn on to measure voltage or current of 200 Hz or less After 30 minutes after warm-up time has passed After measurement range is changed (zero-level compensation) Update interval is 250 ms DC ± (0.
GPM-8310 User Manual External current sensor input (/EX1) 1 mV/°C External current sensor input (/EX2) 50 µV/°C accuracy obtained by doubling the measurement Accuracy when the crest range error for the accuracy when the crest factor is factor is set to 6 or 6A set to 3 When the data update interval is 100 ms, and Auto, Accuracy changes caused add 0.05% of reading to the 0.1 Hz to 1 kHz by data update interval accuracy.
APPENDIX Accuracy of apparent voltage accuracy + current accuracy power S Accuracy of reactive power accuracy of apparent power + (√1.0004 - λ2) - (√1 λ2) ×100 % Q ± [(λ-λ/1.0002)+丨 cosø-cos{ø+sin-1 (influence from the power factor when λ = 0%/100)}丨] ±1 digit Accuracy of power factor λ when voltage and current are at the measurement range rated input ± [丨 ø-cos-1(λ/1.
GPM-8310 User Manual Auto-range decline The range is downed when all of the following conditions are met. Crest factor 3 Urms or Irms is less than or equal to 30% of the measurement range. Urms or Irms is less than or equal to 125% of the next lower measurement range. Upk, Ipk value of the input signal exceeds 300% of the currently set measurement range. Crest factor 6 or 6A Urms or Irms is less than or equal to 30% of the measurement range.
APPENDIX Frequency Voltage Peak Current Peak Active Power Peak Total Harmonic Distortion Mathematical Computation Maximum Current Ratio IHz and VHz V+pk and V-pk I+pk and I-pk P+pk and P-pk THDI and THDV MATH MCR Frequency Measurement Item Measurement item Measurement frequency range Measurement range Frequency filter Accuracy Specifications Voltage and current Data update interval Measurement Frequency Range 20 Hz ≤ f ≤ 100 kHz 0.1 s 10 Hz ≤ f ≤ 100 kHz 0.25 s 5 Hz ≤ f ≤ 100 kHz 0.5 s 2.
GPM-8310 User Manual Integration Item Mode Timer Accuracy Range setting Timer accuracy Remote control Specifications Select manual integration mode, standard integration mode, or repetitive integration mode. Automatically stop integration by setting a timer. Selectable range: 0 hours 00 minutes 00 seconds to 9999 hours 59 minutes 59 seconds ±(Power accuracy (or current accuracy) + 0.1% of reading) (fixed range) Auto range or fixed range is available for Integration ±0.
APPENDIX + 0.35% + 0.35% + 0.50% of of range of range range 440 Hz ≤ f < 0.20% of 0.20% of 0.40% of 1.2kHz reading reading reading + 0.35% + 0.35% + 0.50% of of range of range range * 50Hz/60Hz Compliant IEC61000-4-7 (Update Rate must be ≥ 0.25S). * Harmonic calculation: FFT method in which FFT data length is divided into 2 types: 1024 and 4096. * FFT data length automatically switches in accord with the Frequency and Update Rate of measured signal.
GPM-8310 User Manual Digital IO Signal (Options) Item I/O control output signal I/O level I/O sink current Specifications OUT1, OUT2, OUT3, OUT4 TTL Max 100mA (per/ch) * Q (VAR), S (VA), λ (PF) and Φ (DEG) are originated from the measured values including voltage, current and active power which go through computation process. In respect to distorted signal input, accordingly, the value acquired from other instruments, which employ different methods, may differ from that acquired from GPM-8310 unit.
APPENDIX Status system The diagram below is a description of the status system 239
GPM-8310 User Manual The extended event register receives information about changes in the condition register, which indicates the instrument’s internal condition. The information is the result of edge detection performed by the transition filter. The following table lists the bit definitions for the condition register: Bit Name Decimal Definition 0 Updating 1 The measured data is being updated. UPD changing from 1 to 0 indicates that updating has been completed.
APPENDIX The transition filter parameters detect changes in the specified condition register bits (numeric suffixes 1 to 16) and overwrite the extended event register in the following ways. Condition Definition RISE The specified extended event register bit is set to 1 when the corresponding condition register bit changes from 0 to 1. FALL The specified extended event register bit is set to 1 when the corresponding condition register bit changes from 1 to 0.
GPM-8310 User Manual The following table describes the Status Byte Register. Bit Name Decimal 0 Not Used 1 (Reserved for future use) 1 Not Used 2 (Reserved for future use) 2 Error Queue 4 One or more errors have been stored in the Error Queue. Use STAT:ERR? to read and delete errors. 3 Extended Event 8 One or more bits are set in the Extended Event Register (bits must be enabled, see STAT:EESE). 4 Message Available 16 Data is available in the instrument's output buffer.
APPENDIX Dimensions Units = mm 243
GPM-8310 User Manual Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. declare that the below mentioned product Type of Product: Digital Power Meter Model Number: GPM-8310 are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to EMC (2014/30/EU), LVD (2014/35/EU), WEEE (2012/19/EU) and RoHS (2011/65/EU).
APPENDIX Power measurement Method Direct read method: Directly read the measurement value measured from power measuring instrument. The average power method: Record the actual power value within a settable period of time and then take the average. A settable period of time isn’t less than 10min. The maximum measurement interval is one second. Energy accumulation method: Measure the energy within a settable period of time and then divide it by the time to get the power.
GPM-8310 User Manual Measurement for large current Voltage measurement mode measured from load side (Connect to ammeter externally). The voltage measurement is accurate. The current measurement on load could be larger than the actual one due to leakage current of multi-measurement voltage.
APPENDIX Introduction to IEC-62301 IEC 62301-2011 standard is an international basic standard for measuring standby power consumption of household appliances which is issued by IEEC. It is a standby power consumption measurement method for the various household appliances, power supply, audio and video appliances to comply with. The latest version for this standard is second edition of German standard IEC62301: 2011 (British regulations EN50564: 2011) which is issued on January, 2011.
GPM-8310 User Manual EUP Directive Lot6 specifications Ecodesign directive for energy-using products: The power loss requirement for the products with external power supply such as information devices, consumer electronics product, household appliances, toys, entertainment and sports products and so on in standby and off mode is as below. Mode/Limit Standby Products with time display function. mode Products without time display function. Shutdown mode 248 2010.01 ≦ 2W ≦ 1W ≦ 1W 2013.01 ≦ 1W ≦ 0.5W ≦ 0.
APPENDIX Connection Guide Rear panel Direct connection: I < 1A Method 1 Source CURRENT I Load I V EXT1 VOLTAGE V+ Source 10V MAX V EXT2 20A MAX Method 2 Load V V- I- I+ 600V MAX CAT II 600V Source CURRENT I A 2V MAX Load I V EXT1 VOLTAGE 10V MAX V EXT2 I+ V+ 20A MAX 2V MAX 600V MAX CAT II 600V Source V V- A ILoad 249
GPM-8310 User Manual Direct connection: 1A < I < 20A Method 1 Source Load CURRENT I I V EXT1 VOLTAGE V+ 10V MAX Source V EXT2 V I- 20A MAX Method 2 A 2V MAX I+ 600V MAX CAT II 600V Source Load CURRENT I I V EXT1 VOLTAGE I+ 10V MAX 20A MAX A I- V+ V V- Source V EXT2 2V MAX 600V MAX CAT II 600V Connection with CT/VT Source Load VT CT CURRENT I I V EXT1 VOLTAGE 10V MAX V EXT2 20A MAX 250 Load V- 2V MAX 600V MAX CAT II 600V Load
APPENDIX Connection with EXT1/2 EXT1 Source CURRENT I Load I V EXT1 VOLTAGE 10V MAX V EXT2 20A MAX EXT2 2V MAX 600V MAX CAT II 600V Source CURRENT I Load I V EXT1 VOLTAGE 10V MAX V EXT2 20A MAX Note 2V MAX 600V MAX CAT II 600V The optional GCP-300 accessory is required for EXT1 and EXT2 connection. Refer to your local dealer for more details.