User manual
REMOTE CONTROL
41
VSET<X>?
Description
Returns the output voltage setting.
Response time
Minimum 80ms
Example VSET1?
Returns the CH1 voltage setting
IOUT<X>?
Description
Returns the actual output current.
Response time
Minimum 80ms
Example IOUT1?
Returns the CH1 output current
VOUT<X>?
Description
Returns the actual output voltage.
Panel operation
See page23
Response time
Minimum 80ms
Example VOUT1?
Returns the CH1 output voltage
TRACK<NR1>
Description
Selects the operation mode: independent, tracking
series, or tracking parallel.
Panel operation
See page27
NR1
0: Independent
1: Tracking series
2: Tracking parallel
Response time
Minimum 70ms
Example TRACK0
Selects the independent mode
GPD-3303 Series User Manual
42
BEEP<Boolean>
Description
Turns the beep on or off.
Panel operation
See page21
Response time
Minimum 70ms
Example BEEP1
Turns on the beep
OUT<Boolean>
Description
Turns on or off the output.
Panel operation
See page21
Response time
Minimum 70ms
Example OUT1
Turns on the output
STATUS?
Description
Returns the GPD-3303D or GPD-3303S status.
Response time
Minimum 400ms
Contents
8 bits in the following format
Bit Item Description
0 CH1 0=CC mode, 1=CV mode
1 CH2 0=CC mode, 1=CV mode
2, 3
Tracking 01=Independent, 11=Tracking series,
10=Tracking parallel
4 Beep 0=Off, 1=On
5 N/A N/A
6 Output 0=Off, 1=On
7 N/A N/A